No. It senses through the wall of transparent tubing, reducing the risk of fluid-circuit contamination and additional flow resistance.

FOREACH ABD Air Bubble Detection Module for identifying bubbles, droplets and gas/liquid states in transparent tubing, available in configurations for transparent flexible tubing with outer diameters of 1.6–6.4 mm
The ABD air bubble detector detects bubbles, droplets and gas/liquid states in transparent tubing. It uses non-contact infrared sensing to distinguish gas from liquid by their different optical responses. The sensor does not contact the fluid, avoiding additional contamination and flow resistance.
Designed for fluid-line monitoring in IVD, life science, laboratory automation, environmental monitoring and food testing equipment, ABD supports bubble alarms, droplet detection, reagent-line status monitoring and equipment fault protection. Configurations cover common transparent tubing ODs from 1.6 mm to 6.4 mm; choose the configuration to match the actual tubing OD.
ABD configurations
Select product specifications and save the configuration for your inquiry.
| Product type | Air bubble detector |
|---|---|
| Detection method | Non-contact infrared sensing |
| Detected states | Bubbles / droplets / gas-liquid states |
| Compatible tubing OD | 1.6 / 2.0 / 2.5 / 3.2 / 4.0 / 4.8 / 6.0 / 6.4 mm |
| Compatible tubing materials | Transparent PU / PVC / PTFE / PFA / FEP tubing |
| Detected media | Liquid / Gas |
| Detectable width | Width > 0.8 mm |
| Communication interface | TTL |
| Communication protocol | Modbus RTU |
| Outputs | UART/TTL digital signal; IO analog voltage 0–5V; IO digital alarm signal |
| Operating voltage | 5V DC ±5% |
| Operating current | < 20 mA |
| Response time | Bubble detection: 6 ms; liquid detection: 6 ms |
| Operating temperature | 15℃–45℃ |
| Storage temperature | -25℃–65℃ |
| Dimensions | 35.4 × 19.6 × 29.4 mm |
| Weight | 6.9 g |
1 Selecting a configuration
The ABD detects air bubbles by monitoring changes in light transmitted through transparent tubing. Match the sensor to the tubing OD and optical properties, then verify the output and alarm threshold with the bubbles you need to detect in the installed flow path.
1.1 Tubing fit and detection conditions
Measure the actual tubing OD and select the matching ABD slot. Establish a detection baseline with the specified reagent, tubing and operating flow rate.
| Selection criterion | Specifications and checks |
|---|---|
| Tubing OD | 1.6 · 2.0 · 2.5 · 3.2 · 4.0 · 4.8 · 6.0 · 6.4 mm. Undersize, oversize or incomplete seating changes the optical path. Do not order the detector by tubing ID. |
| Tubing material | Listed transparent tubing: PU/PVC/PTFE/PFA/FEP. Test batch-specific transparency, surface and wall thickness. Do not extend results directly to other materials or rough surfaces. |
| Bubble size | Representative condition: width at least 0.8 mm. This width concerns bubbles spanning the tubing ID; it does not mean every 0.8 mm-diameter microbubble is detectable. |
| Detection location | Secure tubing, clean its surface and control ambient light. Prevent movement in the groove. Remove contamination or attached droplets, then recheck signals. |
2 Installation and integration
2.1 Power supply, analog output and TTL communications
Determine whether your controller will read the analog voltage, the alarm output or a communications register, then wire the corresponding pins. For an RS-485 controller, use an appropriate converter between the module’s TTL interface and the bus.
| Pin | Signal / function |
|---|---|
| 1 | +5 VDC · Supply positive |
| 2 | GND · Supply return / ground |
| 3 | 0–5 V · Photoelectric analog signal |
| 4 | Rx · TTL |
| 5 | Tx · TTL |
| 6 | IO alarm: high level 5 V; pull-up required |
Pin 6: applied voltage <25 V; sink current <50 mA. Check compatibility with the controller input.
Functional schematic. Check the corresponding drawing for mounting dimensions and connector orientation.
1/2: 5 V supply positive/GND. Check 5 V ±5% supply and common ground. Do not apply 24 V.
3: 0–5 V photoelectric analog signal. Check controller ADC voltage limits and compare voltages with actual liquid versus air.
4/5: Rx/Tx at TTL levels. Cross-connect controller Tx/Rx and configure protocol/address. Do not connect directly to standard RS232 or RS485 levels.
6: IO alarm; pull-up required. Design for the output circuit and permitted voltage/current. Do not power a load directly from the output pin.
2.2 Detection threshold, output polarity and alarm logic
Record the signals with liquid and with air in the tube before setting the threshold. Verify the switching threshold and alarm-output polarity separately.
Normal baseline: Liquid generally gives higher values; bubbles lower values. Record the range of signals using the actual tubing, liquid and flow rate. Set and verify the threshold for each reagent used.
Trigger threshold: Set from the actual difference between normal and bubble signals. Test smallest target bubbles, consecutive bubbles and color changes. Check false positives and missed detections.
Output polarity: Check instrument inputs and the Setting configuration. Test bubbles, normal liquid, disconnected wiring and power loss. The instrument must implement fault handling.
3 Mounting, calibration and commissioning
Validate the complete chain: bubble passage, module output, controller reading, instrument alarm and response.
- Clamp and secure the tube: Match OD, fully seat tubing without twist, secure wiring and clean the optical path.
- Check electrical connections: Connect power and required outputs to the drawing. Check pull-ups, common ground and communication levels.
- Establish liquid and air signals: Record normal and target-bubble signals with actual liquid/tubing, then set threshold and polarity.
- Validate the alarm sequence: Test continuous flow, start/stop, different flow speeds, low reservoir level and power-cycle restart. Confirm instrument action.
4 False alarms, missed bubbles and maintenance
Check tubing and optics before thresholds and communication. Do not assume every symptom is a sensor fault.
| Symptom | What to check |
|---|---|
| Continuous alarm | Empty tube, incomplete seating, contamination or threshold. Restore steady liquid and compare analog values. Confirm controller reading and alarm polarity. |
| Target bubble not detected | Bubble span across the tube ID, flow speed and sampling. Check that the bubbles to be detected are within the specified limits. Then verify that the controller’s sampling and filtering retain the relevant signal. |
| Fault after changing tubing or liquid | Transmission, wall thickness, liquid color and baseline change. Record signals and recalibrate; do not directly reuse old thresholds. |
| Communication fault | TTL levels, address, Rx/Tx and CRC. Test one unit’s electrical interface and protocol first. Do not solve faults by arbitrarily increasing voltage. |