How to Choose a Sampling Probe: Diameters, Working Length, Tip and Side Ports

The inner diameter governs pressure drop and internal volume. The outer diameter determines whether the probe fits and is stiff enough. Working length governs access at minimum liquid level and bottom clearance. The tip and side ports determine whether near-bottom aspiration seals the inlet, clogs or leaves liquid behind. These four design choices are linked. Start with the target volume, vessel and motion sequence, then establish the four sets of dimensions. Focusing on one dimension in isolation shifts the problem elsewhere.

This guide is for fluidics designers, mechanical designers and field service personnel working on IVD and laboratory automation instruments. It explains what each dimension controls, then covers inner diameter, outer diameter, working length, tip geometry and side ports, in a sequence you can calculate, check and verify.

01What Do the Four Sampling Probe Dimensions Control?

Distinguish two lengths first. Overall length is the drawing dimension from the probe holder mating face to the tip. Working length is the usable projection from the mounting datum, such as the underside of a flange or the end of a guide sleeve, to the tip. This determines whether the Z-axis travel is sufficient and whether the probe can aspirate at minimum liquid level. The difference is the mounting portion, which includes the flange thickness, bends, guide sleeve and sealing structure.

Wall thickness may be omitted from the order but follows from the diameters: t = (OD − ID) ÷ 2. It affects bending stiffness and the strength margin for welding or bonding at the holder.

DimensionWhat it controlsTypical symptoms when out of range
Outer diameter ODVessel fit, piercing force, bending stiffness and cLLD electrode diameterCannot reach the bottom, rubs the wall or bends on contact with the bottom
Inner diameter IDBore volume, flow velocity, pressure drop and clogging marginAspiration failure, bubble formation or excess residue
Working lengthLiquid level coverage, bottom clearance and Z-axis travelCannot aspirate at minimum level or strikes the bottom
Tip and side portsInlet position near the bottom, clogging risk and dead volumeSealed end face, blocked port or incomplete rinsing

02How to Size the Inner Diameter: Volume and Pressure Drop

Set the inner diameter first because it controls the bore volume, the liquid velocity and the pressure drop. Estimate all three before deciding whether a larger bore is needed to resist clogging.

  • Bore volume: V = π(ID/2)² × L
  • Liquid velocity: v = Q/π(ID/2)²
  • Laminar pressure drop: Δp = 128μLQ/(πID⁴)
The third equation gives the main scaling relationship: pressure drop is inversely proportional to the fourth power of the inner diameter. Halving the bore at the same flow rate increases this pressure drop sixteenfold.
Inner diameter IDLiquid velocityReynolds number ReLaminar pressure drop Δp (100 mm)
0.4 mm7.96 m/s3183159 kPa
0.5 mm5.09 m/s254665 kPa
0.6 mm3.54 m/s212231 kPa

Calculated example, not a product rating. Assumptions: water at 20 °C, μ ≈ 1 mPa·s, probe tube length 100 mm and aspiration flow rate 1 mL/s.

Two limits apply to this table. First, Re reaches the transitional regime and the laminar calculation gives a lower estimate. Entrance losses, side-port restriction, wall roughness and trapped bubbles can increase actual pressure drop. Second, the 159 kPa of suction required in the 0.4 mm case exceeds the available limit for water at 20 °C (about 99 kPa, the difference between atmospheric and vapour pressure). Bubbles form in the probe before the specified flow can be sustained. Increasing suction cannot make up for any further bore reduction in this case.

Allowing more aspiration time is also effective. For the same 0.5 mm bore, extending aspiration from 1 s to 3 s reduces Q to about 0.33 mL/s and reduces the calculated pressure drop in proportion, to about 22 kPa.

Inner-diameter selection sequence:

  • Determine the flow rate: Q = target volume ÷ allowed aspiration time.
  • Calculate the bore from the allowable negative pressure in the probe before selecting a pump for that bore.
  • Check clogging with the bore and side ports together for viscous liquids or media containing fibres, clots or cells. Avoid relying on more suction alone.

03How to Size the Outer Diameter: Vessel Fit, Piercing and Stiffness

The vessel imposes the first limit on outer diameter. The opening of a microtube, conical tube or deep-well plate sets the upper limit, with extra clearance for lateral runout. In a conical vessel, the tip can touch the wall before reaching the intended depth.

The second limit is stiffness. For a uniform cantilever, deflection scales with the cube of the length and inversely with the second moment of area: δ ∝ L³/(OD⁴ − ID⁴). Reducing OD from 1.0 to 0.8 mm with ID held at 0.5 mm reduces the second moment to about 37% and increases deflection to about 2.7 times at the same free length. Increasing ID from 0.5 to 0.6 mm at unchanged OD leaves about 93% of the second moment of area.

Increase OD first for stiffness and ID first for flow, bearing in mind that a larger ID thins the wall. If these requirements conflict, meet vessel fit and stiffness requirements first, then improve inlet flow using side ports.

Third, consider piercing and liquid level detection. A pointed tip is used to pierce film or caps, with sufficient OD to resist bending. In capacitive liquid level detection (cLLD), the probe acts as the electrode. An insulating outer coating affects the capacitive geometry. Specify the coated OD on the drawing because it also changes insertion clearance.

04How to Set Working Length from Minimum Liquid Level and Bottom Clearance

Working length is set by three contributing dimensions:

Lₑff ≥ Δh + s + c

Δh is the change from the highest to the lowest liquid level, including the low-level alarm threshold. s is the working segment from the lowest side port to the tip. c is the safety clearance between the tip and vessel bottom, determined by the bottom geometry, lateral runout and Z-axis positioning repeatability. The overall length includes the mounting portion, such as the flange, bends and guide sleeve.

  1. Mark three heights on the drawing: maximum liquid level, minimum liquid level and vessel bottom.
  2. Set c for the bottom geometry and runout. Flat, conical and U-shaped bottoms differ. A conical bottom may allow a smaller c if lateral runout is also checked.
  3. Add s to obtain the minimum working length.
  4. Add the mounting portion and bends for the overall length.
  5. Check the free length. A 20% increase in L_eff increases deflection by about 73% for the same section (1.2³ ≈ 1.73).

Example: a 316L probe with 0.8 mm OD, 0.5 mm ID and 20 mm free length deflects about 41 µm under a 0.05 N lateral load. With a 1.0 mm OD in the same conditions, deflection is about 15 µm. Increasing the original probe's free length to 24 mm raises the 41 µm value to about 71 µm.

Bend radius

Under the room-temperature bending requirements cited in JB/T 5000.11, for an OD no greater than 42 mm the radius between the two probe tube segments should be at least 2.5 times the OD, with a tolerance of ±3 mm. Using this calculation, 0.8 mm OD gives a 2.0 mm minimum radius and 1.2 mm OD gives 3.0 mm.

Bending changes both the working length and the direction of the tip. Specify developed length, bend radius and bend direction on the drawing. A tight bend can flatten the bore and leave areas that trap residue or are difficult to rinse. In practice, thin-walled tubes are often bent to 3–4 times the OD, with the tip position rechecked after assembly.

05Sampling Probe Tips: Pointed, Flat and V-Notched

Tip geometry determines where liquid enters and where the force is applied when the probe tip contacts the vessel bottom.

  • Pointed: used for piercing film or caps. Stress concentrates at the point, which is vulnerable to bottom strikes. Evaluate piercing force, landing position and membrane thickness together.
  • Flat: a planar end simplifies tool setting and drawing callouts. Contact with a flat bottom can form a sealing liquid film. A side port provides an inlet in this configuration.
  • V-notched: the V-shaped cross-section is less prone to blockage at a single point and suits viscous or particle-bearing media. The tip is less stiff, so assess the risk of striking the bottom.

06Side Ports: Diameter, Position and Residual Liquid

Side ports keep an inlet open when the end face seals against the bottom, redirect intake laterally to reduce disturbance of sediment and cells, and provide a bypass if one opening clogs.

The sizing criterion is that total side-port area is at least the cross-sectional area of the probe bore.

n × (πdₕ²/4) ≥ πID²/4 ⇒ dₕ ≥ ID/√n

Two ports require d_h ≥ 0.71 ID, and four ports require d_h ≥ 0.50 ID. Undersized ports restrict flow and can cause higher suction, bubble formation and intermittent aspiration. This can be mistaken for insufficient pump capacity.

Residual liquid governs the port position. The segment of height h from the lowest port to the tip is dead volume: V_dead = π(ID/2)² × h. With a 0.5 mm bore and h = 2 mm, this is about 0.39 µL, or 2% of a 20 µL dispense and 8% of a 5 µL dispense.

Orient ports away from the vessel wall and sediment, typically upward or diagonally upward. More ports spread the flow across openings and make rinsing harder. Use the fewest needed to meet the total area criterion.

Changes to the ports should be carried through to the wash station. Port retention, internal surface area and rinsing coverage all change. Determine whether residue is acceptable by measuring carryover after cleaning.

07Frequently Asked Questions

Are standard sampling probe models available?

Sampling probes are mainly made to instrument drawings. Confirm OD, ID, length, tip and mounting against the drawing. The website describes custom options rather than a standard specification table.

Does a smaller bore aspirate more cleanly?

A smaller bore increases velocity, shear and pressure drop. Shear can be a source of error for samples containing cells or macromolecules. Residue also depends on dead volume, wall roughness and washing.

Can I increase working length if the probe already hits the bottom?

Determine whether the probe is too long or has too little bottom clearance. Increasing length increases free projection and deflection, which scales with its cube, and can lead to rubbing at the wall. Check c and positioning repeatability before changing the probe or the motion.

Are more side ports better?

Ports provide the required total area and a bypass in case of blockage. Too many distribute flow between ports and increase retention and washing difficulty. Use the minimum count that meets the area criterion.

Can a coated probe be swapped directly into a cLLD system?

The coating changes OD and capacitive geometry and can shift the triggering point. Recalibrate level detection after replacement and confirm that the coated probe still fits the vessel.

What should the drawing specify?

Include OD, ID, overall length, working length, mounting datum, bend direction and angle, tip geometry, side-port position and diameter, coated OD and the position of the tip relative to the vessel bottom.