What Is a Compression-Ring Fitting? Design, Sizes and Selection

A compression-ring fitting is a flangeless threaded connection for instrument fluidics. It consists of a fitting nut, a compression ring and a sealing ferrule. After square-cutting the tubing, the matched ring and ferrule are preassembled on the end. The nut then loads this assembly against the receiving port to make the seal. Applications include reagent transfer, sample handling and wash lines linking tubing to pumps, valves and fluidic modules in analytical instruments and laboratory automation.

FOREACH HSF6 compact fittings are available with M6×1 and 1/4-28 UNF threads. The configurations covered here accept 1.6, 2.0, 2.5, 3.0 and 3.2 mm OD tubing. Nut materials include PPS and PEEK; the CL-1S kit combines a stainless-steel compression ring with a PEEK sealing ferrule. The illustrated HSF6 has a 6.0 mm head diameter and a 20.0 mm overall length. Use these dimensions to assess port spacing and axial clearance. Thread, tubing-size and material combinations are listed below.

These connections support on-site assembly and prefabricated tubing assemblies with specified lengths, end fittings and identification labels. Complete tubing ends can be disconnected and reconnected for maintenance as specified. This article focuses on HSF6/CL-1S and separately describes CL-1X-16 for smaller 6-40 UNF connections. Select other series and sizes against their own drawings.

01What is a compression-ring fitting? Nut, ring and ferrule

Here, compression-ring fitting means a microfluidic tubing connection using a separate compression ring and sealing ferrule. It is a type of flangeless connection. Flangeless means no preformed tubing flange is required; it does not imply that all such fittings use the same parts. The compression ring is not a circlip or snap ring used in a shaft groove.
ComponentFunctionSelection criteria
Threaded fitting nutScrews into the instrument port and applies axial load to the tubing-end kitThread, tubing clearance, nut material and external dimensions
Compression ringFits around the mating portion of the ferrule to compress and retain the tubing-end assemblySelect with the matching ferrule; check orientation and preassembly requirements
Sealing ferruleMates with the tubing and port to form the front sealing interfaceTubing OD, material, end face and position
Tubing and receiving portForm the flow path and mating sealing surfacesTubing ID/OD, cut quality, port bottom and installation depth

In CL-1S, the compression ring is SUS stainless steel and the sealing ferrule is PEEK. The nut, ring and ferrule are separate components. A PEEK or PPS nut does not mean all wetted surfaces in the connection use that material.

02How does it seal, and why use prefabricated tubing?

As the nut screws into a flat-bottom receiving port, it transfers axial force to the ring and ferrule. Compression at the specified position brings the tubing-end assembly against the port sealing surface. Both ring-to-ferrule fit and tubing-end position matter. The thread provides retention and load; correct contact at the sealing interface must be checked.

Check that the tubing end is flush with the ferrule face

Tubing-end positionPossible effectAssembly check
Flush with the ferrule faceProvides the specified end positionSquare, clean cut; position maintained after pressing
Recessed inside the ferruleMay leave extra space at the connectionCheck cut length and tubing movement during preassembly
Protruding or cut at an angleMay contact the port prematurely and interfere with sealingRecheck end preparation and assembly position

The photograph below identifies the fitting nut, tubing, stainless-steel compression ring and sealing ferrule, showing how the components fit together.

Prepare the tubing end before connecting the instrument

Specify tubing, length, end components and labels before prefabrication. Completing the ends first supports batch assembly and line identification. During connection, the nut can rotate relative to the tubing, reducing twist caused by tightening. This is not a rotary union for continuous rotation during operation.

A serviceable complete tubing-end assembly can be retained during maintenance, but inspect the ferrule face, ring position and tubing condition. Reconnecting an intact end is different from removing an old ring and fitting it to new tubing.

03HSF6 threads, tubing sizes, materials and dimensions

HSF6 identifies the fitting nut series; CL-1S and CL-1X identify the ring-and-ferrule kits. Select the nut for the port, then the kit for the actual tubing OD. These three series are not interchangeable fitting models. The table covers the HSF6 configurations discussed here, not every compression-ring product.
Complete fitting modelThreadCompatible tubing ODNut material
HSF6-M6-16-PS-BM6×11.6 mmPPS
HSF6-M6-32-PS-BM6×12.0, 2.5, 3.0, 3.2 mmPPS
HSF6-U28-16-PK-N1/4-28 UNF1.6 mmPEEK
HSF6-U28-16-PS-B1/4-28 UNF1.6 mmPPS
HSF6-U28-32-PK-N1/4-28 UNF2.0, 2.5, 3.0, 3.2 mmPEEK
HSF6-U28-32-PS-B1/4-28 UNF2.0, 2.5, 3.0, 3.2 mmPPS

One nut size can require several different tubing kits

Actual tubing ODHSF6 nut size codeRing-and-ferrule kitKit materials
1.6 mmSize 16, M6/U28CL-1S-16SUS ring + PEEK ferrule
2.0 mmSize 32, M6/U28CL-1S-20SUS ring + PEEK ferrule
2.5 mmSize 32, M6/U28CL-1S-25SUS ring + PEEK ferrule
3.0 mmSize 32, M6/U28CL-1S-30SUS ring + PEEK ferrule
3.2 mmSize 32, M6/U28CL-1S-32SUS ring + PEEK ferrule

For 2.5 mm OD tubing in an M6 flat-bottom port, start by checking HSF6-M6-32-PS-B with CL-1S-25. The code 32 identifies this nut size group. The kit must still be size 25 for 2.5 mm tubing; do not choose CL-1S-32 simply to match the nut code.

CL-1X-16 is a separate small-port configuration

CL-1X-16 accepts 1.6 mm OD tubing and uses a SUS ring with a PEEK ferrule. Its matching fitting is HFL6-U40-16 with a 6-40 UNF thread. Although it accepts the same tubing OD as CL-1S-16, the port, kit geometry and fitting differ; they are not directly interchangeable.

External dimensions determine clearance; installation depth also depends on the port

Illustrated HSF6 dimensionValueUse during installation
Head diameter6.0 mmCheck adjacent fitting spacing and access for tightening
Overall nut length20.0 mmCombine with installation depth to determine exposed length
Axial dimension of the front section10.5 mmCheck with port depth and kit position; not effective thread engagement or actual installation depth

These dimensions describe the illustrated HSF6, not HSF, HFL6 or other configurations. The assembly must reach its sealing position while retaining adequate thread engagement. If the nut bottoms against the instrument surface before the ferrule seats, additional torque will not correct the dimensional mismatch.

04Choosing between compression-ring, ferrule and flanged fittings

Compare tubing-end preparation and sealing components first, then production assembly and maintenance. Compression-ring and conventional ferrule connections both avoid flanging. The former uses a separate ring and sealing ferrule; the latter uses a matched ferrule to grip the tubing and participate in sealing.
ComparisonCompression-ring connectionConventional flangeless ferrule connectionFlanged connection
End preparationSquare-cut, then install ring and ferruleSquare-cut, then fit the ferrulePreform a tubing flange
Main componentsNut, compression ring and sealing ferruleNut and ferruleNut, tubing flange and matching washer
Assembly focusPreassemble with dedicated pliers; check orientation, flush end position and pressingCorrect ferrule, orientation and tubing seatingFlange quality and sealing-face contact
MaintenanceInspect the complete tubing end before reconnecting as specifiedPrepare replacement ferrules for the single-installation types covered in the companion articleInspect flange, washer and port before reconnecting
Situations to evaluate firstPrefabricated tubing and maintenance retaining the assembled endFlangeless assembly with accessible ferrule replacementEstablished flanging process or prefabricated flanged assemblies

These are assembly differences, not pressure rankings. For frequent maintenance, compare inspection and replacement requirements. For closely spaced ports, compare nut dimensions and tool access instead of selecting by connection name alone.

  1. Flangeless ferrule fittings: design, sizes and selection
  2. Flanged tubing assemblies: sealing, port compatibility and selection

05Applications and tubing-assembly configuration

Begin with assembly and maintenance tasks when evaluating compression-ring connections. These examples illustrate selection considerations; suitability also depends on the fluid, temperature and pressure.
ApplicationPurpose of a tubing assemblyConfiguration priorities
Reagent transfer in analytical instrumentsPrepare connections and lengths before instrument installationEnd ports, wetted materials, tubing ID and length
Sample handling and wash linesControl end position and simplify line identification and servicingEnd gaps, cleaning conditions, carryover requirements and labels
Connections between pumps, valves and fluidic modulesStandardize assembly and spare partsPort spacing, tubing direction, pressure pulsation and support
Modules removed for maintenanceRetain and reconnect the complete tubing-end assemblyExpected reconnection frequency, sealing-face inspection and replacement rules

FOREACH assemblies can use FEP or PTFE tubing with specified lengths and labels. OD determines nut and kit compatibility; ID and length affect flow resistance and internal volume. Tubing with the same OD but different IDs can behave differently in flow and after pressing.

For selection, provide both port drawings, tubing ID/OD, finished length and measurement reference, fluid and cleaning conditions, temperature, continuous and peak pressures, and maintenance requirements. Select each end separately if the ports differ. O-rings or stop sleeves on the tubing limit nut travel; they are not the instrument-port seal.

06Installation, pressure capability and tightening

Separate initial preassembly from connection to the port

Initial assembly requires the dedicated compatible pliers to press the compression ring and sealing ferrule into position. Hand pressure alone is unlikely to assemble them correctly. Tightening the nut or increasing port torque cannot replace preassembly. For maintenance reconnection, inspect the already assembled tubing end before inserting it into the port.

FOREACH assembly pliers are model T-SF-4-SS, product code 959016. When assembling tubing yourself, obtain the tool compatible with the kit and tubing size and confirm clamping position and operating instructions.

  1. Cut the tubing square: use a suitable cutter, remove debris and check for flattening, burrs or cracks.
  2. Check orientation: slide on the nut, ring and ferrule in order, following the selected kit diagram.
  3. Position and preassemble: align the tubing end flush with the ferrule face, press with the dedicated compatible assembly pliers, then check ring seating and tubing recession or protrusion.
  4. Connect the port: clean and inspect the sealing surface, hold the assembly in position and tighten as specified.
  5. Check the connection: secure the tubing and test for leakage, air ingress and end movement at the specified fluid, temperature and pressure conditions.

What pressure can the fitting withstand?

Pressure capability applies to the complete nut, kit, tubing and port combination at stated temperatures and operating conditions. No common HSF6/CL-1S pressure rating is assigned here. Obtain the allowable working pressure for the selected combination. Check positive pressure, vacuum and start/stop peaks separately; a PEEK nut or metal ring alone does not establish pressure capability.

What installation torque is recommended?

FOREACH plastic rigid-tubing fittings with M6×1 or 1/4-28 UNF threads have a recommended installation torque of 0.25–0.30 N·m. Plastic U40 compression-ring fittings with 6-40 UNF threads use 0.08–0.15 N·m. Match the recommendation to the thread and nut material.
Thread sizeApplicable fittingRecommended installation torque
M6×1Plastic rigid-tubing fitting0.25–0.30 N·m
1/4-28 UNFPlastic rigid-tubing fitting0.25–0.30 N·m
6-40 UNFPlastic U40 compression-ring fitting0.08–0.15 N·m

These values apply to threaded tightening into the port, not the pressing force for kit preassembly. Preassemble the ring and ferrule correctly, tighten to the corresponding torque, then check sealing and tubing-end position.

Reconnection and troubleshooting

SymptomCheck firstAction
Leakage after first connectionMatched parts, pressing orientation, flush tubing end and port bottomStop and depressurize; correct positioning or component problems and retest
Leakage after maintenance reconnectionDirty or damaged ferrule face, ring movement or tubing deformationDetermine whether cleaning, repositioning or assembly replacement is needed
Reduced flow after assembly replacementTubing ID, length, debris, end compression and tight bendsCompare flow and pressure drop with the same fluid and pump conditions
Passes static check but fails during operationStart/stop peaks, temperature changes, vibration and external loadsCheck support and movement allowance; validate under actual operating cycles

Reconnection as specified does not mean unlimited cycles or maintenance-free use. For frequent disconnection, define test cycles, operating conditions and allowable leakage. Record sealing, tubing-end position and flow after each cycle. Address worn faces, cracks or ring movement according to the assembly requirements.

07Frequently asked questions

Is a compression-ring fitting also flangeless?

Yes. The ring-and-ferrule design described here uses square-cut tubing without a preformed flange. Conventional ferrule fittings can also be flangeless, but parts and assembly differ; the name alone does not establish interchangeability.

If the HSF6 model has size code 32, must the kit also be size 32?

No. The size-32 HSF6 nuts listed here cover tubing ODs of 2.0, 2.5, 3.0 and 3.2 mm, while CL-1S is selected for the actual OD. For example, 2.5 mm tubing needs CL-1S-25. Nut size group and kit tubing size are separate selections.

Are CL-1S-16 and CL-1X-16 interchangeable?

No. Both accept 1.6 mm OD tubing, but CL-1S corresponds to the HSF6 M6/U28 configurations here, while CL-1X-16 pairs with HFL6-U40-16 for a 6-40 UNF connection. Check nut, port and kit geometry together.

Is a dedicated assembly tool required?

Yes. Initial kit preassembly requires compatible dedicated pliers: FOREACH T-SF-4-SS, product code 959016. Hand pressure alone is unlikely to seat the ring and ferrule correctly, and higher nut torque cannot replace preassembly. Check tool compatibility with the kit and tubing size. Inspect complete, previously assembled ends before reconnecting.

What torque should be used for installation?

FOREACH plastic rigid-tubing fittings with M6×1 or 1/4-28 UNF threads use 0.25–0.30 N·m; plastic U40 compression-ring fittings with 6-40 UNF threads use 0.08–0.15 N·m. These recommendations apply to threaded port tightening, not kit preassembly force.

Does reconnection mean an old ring can be transferred to new tubing?

Not necessarily. Inspect sealing surfaces, ring position and tubing condition before reconnecting an intact end. Removing an old ring and installing it on another tube is a separate operation; permission depends on the specific kit requirements.

Does a flush tubing end guarantee zero dead volume?

No. A square cut and correct end position help control gaps, but residual volume also depends on port geometry, tubing ID and assembly condition. Assess sample and wash lines against carryover and cleaning requirements.

Need help selecting fittings or prefabricated tubing?

Provide port drawings, tubing ID/OD, finished length, fluid, temperature, pressure and maintenance requirements to confirm matched parts, installation and sample validation.