Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control Series
Applications
IVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Fluidic Components for
Analytical Instruments

Pumps, valves, probes, fittings, tubing and sensing components for precise sample preparation, reagent handling and instrument fluidics.

Key Fluidic Tasks

  • Sample aspiration and metering
  • Reagent delivery and path switching
  • Rinsing, drainage and waste handling
  • Pressure, liquid-level and bubble monitoring
Home/Applications/Analytical Instruments
  • Chromatography / Autosampler
  • Spectroscopy / Elemental Analysis
  • Water Quality / Environmental Monitoring
  • Sample Preparation Systems
  • Laboratory Analyzer System Integration

Fluidic Valves for Detector Diversion, Bypass and Waste

Application guide · Diversion

A diversion valve sends liquid to detector, reaction path, collection or waste according to the analytical window. Its value is the state transition: the controller must know when the required composition has reached the valve and where any off-spec leading or trailing volume will go.

Engineering example: a 200 μL path between a mixing junction and diversion valve operated at 1 mL/min introduces about 12 seconds of nominal transport delay. Switching at the mixer command time would therefore send the wrong composition to the detector unless transit and dispersion are included.[1][2]

Application guides

Analytical instruments

FOREACH in analytical instruments
Piston pumps in analytical instruments
Sample aspiration and transferReagent and standard dispensingDilution, spiking and proportioningIncremental titrant dosing
Syringe pumps in analytical instruments
Automated sampling and injectionMultichannel reagent distributionAutomated dilution and standardsSequential injection and flow analysis
Pipetting pumps in analytical instruments
Disposable-tip sample transferTip-based dilution and distributionSample-preparation workflows
Valveless metering pumps in analytical instruments
Repeated reagent dispensingTitrant meteringCarrier and flow-cell supplyProportional liquid preparation
Diaphragm pumps in analytical instruments
Wash-liquid delivery and primingWater and working-liquid replenishmentThermal and auxiliary circulationGas-liquid waste aspiration
Fluidic valves in analytical instruments
Multi-source flow selectionSample-loop injection and high-pressure switchingDetector diversion, bypass and wasteTimed shutoff and isolation
Probes and mixing in analytical instruments
Sampling and liquid deliveryProbe-path washingReaction-vessel and microplate washingReaction mixing
Fluid-state monitoring in analytical instruments
Bubble and liquid-state detectionPressure and blockage monitoring
Tubing and connections in analytical instruments
Tubing material and dimensionsFittings and assemblyFiltration and backflow protection

ON THIS PAGE

Define the liquid, receiver and complete duty before selecting the componentCalculate the working point and the margin the instrument really needsMap the task to a FOREACH configuration and the rest of the fluid pathProgram the full liquid cycle, including prime, wash and recoveryDesign abnormal-state diagnosis before the fault occursAccept performance where the liquid is actually usedReferencesRelated guides

01Define the liquid, receiver and complete duty before selecting the component

Translate the analytical task into an engineering duty
Use caseInputs to defineDecision supported
Detector bypassStabilization time and valid read windowKeep prime and transition liquid away from the detector
Fraction or sample collectionArrival window and collection volumeCoordinate valve state with actual transit
Waste isolationCleaning chemistry and safe destinationPrevent cleaner from reaching sensitive components
Keep published examples inside their boundary

The referenced international-manufacturer workflow explains how a mature instrument organizes this task. It does not transfer that manufacturer's specifications to a FOREACH component. FOREACH capability is stated only through the cited product information and must be confirmed at the selected working point.

02Calculate the working point and the margin the instrument really needs

Calculations to complete before prototype selection
QuantityHow to define itWhy it matters
Nominal delayupstream volume ÷ installed flowStarting point for switch timing
Transition banddispersion and mixing around concentration changeDefines waste allowance
Outlet balanceresistance of each destination pathPrevents flow change when state changes

Use minimum, routine and maximum conditions rather than one nominal point. Include the first cycle after priming or idle, the lowest source level, the highest expected restriction and the actual receiving geometry. A nominal capacity, free-flow value or theoretical command increment is not an acceptance result.

03Map the task to a FOREACH configuration and the rest of the fluid path

Configuration routes to compare[2]
ConfigurationWhen it is a useful starting pointBoundary to confirm
Rotary diversion valveSeveral destinations or method statesMap ports and common swept volume
Solenoid switching valveFast two-path diversionConfirm normal state and pressure balance
Feedback or sensor triggerVariable arrival timeSwitch from measured event when suitable
  • FOREACH switching valves
  • Discuss this analytical-instrument duty with FOREACH

04Program the full liquid cycle, including prime, wash and recovery

Prime to waste→Establish stable composition→Wait for transit or sensor trigger→Switch to valid destination→Complete acquisition→Divert wash and return safe
A complete method cycle is more than the active delivery step

The exact physical order and parallel actions depend on the instrument. Retain each state in the control and verification plan.

Record valve states, motion, dwell, sensor windows and rejected recovery cycles. This makes throughput and liquid consumption auditable and prevents a fast component movement from being reported as the complete analyzer cycle.

05Design abnormal-state diagnosis before the fault occurs

Observed result, likely mechanism and useful next action
Observed conditionLikely mechanismEngineering response
Detector sees wash frontSwitch too early or wrong stateMeasure actual transit with tracer
Flow changes on diversionUnequal branch resistanceBalance outlets or control flow independently
Collected fraction is dilutedBroad transition or excess common volumeReduce volume and narrow collection window

The controller should distinguish a recoverable event from a result-invalidating event. A retry is useful only after the original cause has been removed and the liquid path has returned to a verified state.

06Accept performance where the liquid is actually used

Minimum qualification plan for the integrated channel
ResultWhere or how to measureAcceptance question
Switch timingTracer transition at final destinationValid window is captured
Branch flowMeasure both states at installed loadNo unacceptable method disturbance
Safe statePower loss and controller resetLiquid reaches the defined safe destination
  1. Test the exact liquid and complete wetted path at minimum, routine and maximum conditions.
  2. Separate first-cycle, steady-state and post-idle results; do not average away recovery behavior.
  3. Challenge the relevant fault and confirm the controller prevents an invalid result from being accepted.
  4. Retain the configuration, software sequence, consumables and receiving-position result together as the qualification record.

References

View references (2)
  1. [1]VICI: Sample Injection with a Cheminert 6-Port Valve
  2. [2]FOREACH: Fluidic Valve Product Range

Related guides

  • Multi-source flow selection
  • Sequential injection and flow analysis
  • Bubble and liquid-state detection

Need help with an analytical instrument fluid path?

Share the medium, flow range, pressure, connection and control requirements with our engineering team.

Contact an Engineer