MEDICA 2026Foreach Booth 1H39-1Medical Trade Fair in Düsseldorf, GermanyDüsseldorfNov 16–19
Learn MoreBook a Meeting
Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
Frequently Asked QuestionsDatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Frequently Asked Questions
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/Analytical instruments
< Back

Diaphragm pumps: select separately for washing, replenishment, circulation and waste removal

Application guide

Diaphragm pumps can supply wash liquid, replenish working liquid, provide auxiliary circulation or aspirate waste in instruments. Liquid-only supply and gas-liquid aspiration require different medium compatibility and inlet conditions; one flow value cannot describe all duties.

Application guides

Analytical instruments

Analytical instruments Overview
Liquid chromatography systems: locate the fluidic task before selecting pumps and valves
Spectroscopy and elemental analysis
Water analysis units
Analytical Sample Preparation
Analytical system integration
Piston Pumps in Analytical Instruments
Automated dilution with piston pumpsPiston Pump Titration
Syringe pump
Syringe-pump samplingMultiple dispensing with syringe pumpsSyringe-pump standard preparation and over-range sample dilutionSequential injection and flow analysis
Air-displacement pipetting pump
Disposable-Tip Sample TransferTip-based dilution and dispensingTip-based sample preparation
Valveless metering pumps
Repeated Dispensing with Valveless PumpsValveless-pump titrationCarrier-liquid and flow-cell supplyTwo-stream proportioning
Diaphragm pump
Wash-liquid supply and primingWorking-liquid replenishmentAuxiliary liquid circulationGas-liquid waste aspiration
Valves for Analytical Fluidics
MRV3 multi-source selectionHigh-pressure sample-loop injectionDetector Flow Diversion and BypassSolenoid-valve shutoff
Sampling, Washing and Mixing Components
Sampling needlesProbe-wash stationReaction-cup and microplate washingReaction mixing
Fluidic monitoring
Bubble and Liquid-State DetectionPressure and Blockage Monitoring
Tubing, fittings, filters and check valves
Tubing materials and dimensionsLow-retention connectionsFiltration and backflow protection

ON THIS PAGE

Locate the source, drive and receiving position for this taskFlow, suction lift and duty cycle describe the installed operating pointCompare product options against task requirementsPlan supply and gas-liquid aspiration separatelyAssess the complete cycle through branch-specific resultsAfter an abnormality or maintenance, first identify which segment needs recoveryContinue with tasks on the same branchReferencesRelated guides

01Locate the source, drive and receiving position for this task

Reservoir or gas/liquid inlet→Diaphragm pump→Actual branch resistance→Supply outlet or waste vessel
Functional or operational sequence; confirm connections and branches for each state on the instrument

Identify whether each task is discrete dosing, continuous flow or drainage, and specify its medium, load and timing. Trace contact surfaces, pressure and destinations along the diagram before assigning pump and connection roles.

02Flow, suction lift and duty cycle describe the installed operating point

Specify actual flow, pressure or suction lift, formulation, temperature and duty cycle. Free flow supports initial screening; the combined load of tubing, valves, wash heads and filters determines the installed operating point.

Compare configuration conditions by actual task
Condition to compareInformation requiredSelection affected
Inlet mediumLiquid, intermittent air-containing liquid or gasSeparate liquid supply from gas-liquid aspiration
Installed operating pointSuction lift, outlet load and target flowDo not use free-flow specifications alone to assess an instrument
Duty cycleStart-stop cycles, duty cycle, temperature and maintenance intervalCheck long-duration operation and recovery

03Compare product options against task requirements

The following candidate options match this page's responsibilities. Compare them against the operating conditions described, then check capacity, materials and interfaces on the product pages. Card images show representative configurations; suitability must still be confirmed at actual operating points and by application results.

Move from task conditions to product configuration
Product categoryCorresponding roleConfirm before selection
DPL30 liquid diaphragm pumpLiquid supply or auxiliary circulationSelect by installed resistance, actual wash or working fluid and task window
DPL60 liquid diaphragm pumpLiquid supply or auxiliary circulationJudge by peak demand and installed operating point, not free flow alone
DPL30H high-backpressure liquid diaphragm pumpLiquid branches requiring higher backpressure capabilityCalculate pressure budget and actual flow first, then confirm interfaces and materials
DPGL800 gas–liquid diaphragm pumpSuitable gas–liquid aspiration or container-vacuum branchesConfirm inlet medium and waste architecture; not for continuous liquid-only transfer (100% liquid)
  • DPL30 liquid diaphragm pump
  • DPL60 liquid diaphragm pump
  • DPL30H high-backpressure liquid diaphragm pump
  • DPGL800 gas–liquid diaphragm pump

04Plan supply and gas-liquid aspiration separately

Define start, stop and source states separately for supply and waste removal. Air entry, foam and depleted sources can occur at different stages. Confirm unloaded or dry-running capability for the specific model.

Action records should link commands, valve or interface states and actual received results. When raw material, container or sample identity changes, reconfirm shared paths and conditions permitting the next step. Completion of drive movement alone does not replace task acceptance.

05Assess the complete cycle through branch-specific results

Measure supply, drainage, peak load and duty cycle under actual installation conditions. Washing requires residue-recovery validation; circulation requires temperature or method targets. Pump rotation alone demonstrates neither.

Link operating-point tests to interpretable results
Validation scenarioMeasurement or observationInterpretation and action
Branch functionMeasure supply, replenishment or drainage results separatelyEvaluate the pump against the actual task
Boundary loadsInclude changes in source level and path resistanceConfirm capacity changes and restart procedures

Record test medium, configuration, installed path and operating sequence; present routine operating points and boundary results separately. Calculations support design discussions, component documentation checks models, and installed measurements confirm the actual task.

06After an abnormality or maintenance, first identify which segment needs recovery

Confusing capacity, free flow and gas flow can lead to an unsuitable drive configuration. Check actual results from the affected step before resuming the next action.

Retain current source, container identity, completed actions, valve or interface state and actual receiving information. Use the table to locate the stage that changed before deciding to retry, continue or restart. After recovery, recheck the relevant operating point before returning an uncertain action to the automatic cycle.

  • Bring the fluidic diagram and operating-point records to discuss configuration

07Continue with tasks on the same branch

  • Wash-Liquid Supply and Priming: Calculating Installed Flow from Wash Volume
  • Working-liquid replenishment: reservoir demand, level hysteresis and recovery
  • Auxiliary liquid circulation: flow, temperature and pulsation tolerance
  • Gas-liquid waste aspiration: direct pumping or headspace vacuum

References

View references (2)
  1. [1]KNF: Pumps for Clinical Diagnostics and Analytics
  2. [2]FOREACH: Miniature Diaphragm Pump Product Range

Related guides

  • Wash-Liquid Supply and Priming: Calculating Installed Flow from Wash Volume
  • Working-liquid replenishment: reservoir demand, level hysteresis and recovery
  • Auxiliary liquid circulation: flow, temperature and pulsation tolerance
  • Gas-liquid waste aspiration: direct pumping or headspace vacuum

Related Products

DPL30 liquid diaphragm pump; representative configuration. Select capacity, materials and interfaces for the application.

DPL30 liquid diaphragm pump

DPL60 liquid diaphragm pump; representative configuration. Select capacity, materials and interfaces for the application.

DPL60 liquid diaphragm pump

DPL30H high-backpressure liquid diaphragm pump; representative configuration. Select capacity, materials and interfaces for the application.

DPL30H high-backpressure liquid diaphragm pump

DPGL800 gas–liquid diaphragm pump; representative configuration. Select capacity, materials and interfaces for the application.

DPGL800 gas–liquid diaphragm pump

Related Technical Articles

How to Read a Diaphragm Pump Flow-Pressure Curve—and Predict Installed Flow

A catalogue value such as 300 or 600 mL/min is not a fixed flow rate inside an instrument. The useful engineering question is: at the actual inlet pressure, outlet backpressure, fluid, tubing and supply voltage, where will the pump and the fluidic circuit settle?

2026-07-28View Details

Selecting a Self-Priming Miniature Liquid Diaphragm Pump: Lift, Priming Time and Dry or Wet Starts

Select a self-priming miniature liquid diaphragm pump by checking suction lift, time to first liquid, dry and wet starting conditions, and stable delivery. Use DPL30, DPL60 and DPL30H specifications to define an instrument-level test.

2026-09-07View Details

IVD Waste Aspiration: Liquid Pump or Vacuum Pump? Direct vs Indirect Aspiration

Probe washing, reaction-cup cleaning, and routine waste removal in IVD instruments are all described as “waste aspiration.” Yet whether waste passes directly through a liquid pump or first enters a collection bottle while a vacuum pump creates negative pressure leads to entirely different selection logic.

2026-08-09View Details

Discuss your diaphragm-pump fluidics

Provide a fluid-path diagram, actual liquids and pressures in each state. Specify liquid, intermittent air-liquid mixture or gas, suction lift, outlet load, target flow, start/stop timing, duty cycle, temperature and maintenance intervals. Compare configurations against the guide's validation objectives.

Contact an engineer