Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/Analytical instruments
< Back

Filter and check-valve protection: analytes, resistance and maintenance limits

Application guide

Filters retain specified particles; check valves limit reverse flow under defined conditions. Protective components also change path volume and resistance, and sample filtration may alter the measured fraction. Define the purpose first.

Application guides

Analytical instruments

Analytical instruments Overview
Liquid chromatography systems: locate the fluidic task before selecting pumps and valves
Spectroscopy and elemental analysis
Water analysis units
Analytical Sample Preparation
Analytical system integration
Piston Pumps in Analytical Instruments
Automated dilution with piston pumpsPiston Pump Titration
Syringe pump
Syringe-pump samplingMultiple dispensing with syringe pumpsSyringe-pump standard preparation and over-range sample dilutionSequential injection and flow analysis
Air-displacement pipetting pump
Disposable-Tip Sample TransferTip-based dilution and dispensingTip-based sample preparation
Valveless metering pumps
Repeated Dispensing with Valveless PumpsValveless-pump titrationCarrier-liquid and flow-cell supplyTwo-stream proportioning
Diaphragm pump
Wash-liquid supply and primingWorking-liquid replenishmentAuxiliary liquid circulationGas-liquid waste aspiration
Valves for Analytical Fluidics
MRV3 multi-source selectionHigh-pressure sample-loop injectionDetector Flow Diversion and BypassSolenoid-valve shutoff
Sampling, Washing and Mixing Components
Sampling needlesProbe-wash stationReaction-cup and microplate washingReaction mixing
Fluidic monitoring
Bubble and Liquid-State DetectionPressure and Blockage Monitoring
Tubing, fittings, filters and check valves
Tubing materials and dimensionsLow-retention connectionsFiltration and backflow protection

ON THIS PAGE

Locate the source, drive and receiving position for this taskConfirm purpose, analyte and added resistance togetherCompare product options against task requirementsPosition protection according to the source and document maintenanceTest pressure drop, backflow and analytesAfter an abnormality or maintenance, first identify which segment needs recoveryContinue with tasks on the same branchReferencesRelated guides

01Locate the source, drive and receiving position for this task

Particle or reverse-flow source→Filter or check valve→Added resistance and retention→Protected component or analytical unit
Functional or operational sequence; confirm connections and branches for each state on the instrument

Specify protection purpose, target, filter material and pore size, check-valve opening requirements, added pressure drop and permitted maintenance. Trace contact surfaces, pressure and destinations before assigning pump and connection roles.

02Confirm purpose, analyte and added resistance together

Specify particle sources, analytes, pore size, filter material, opening pressure and actual load. Check-valve action is not sterile isolation, and filter protection does not automatically comply with an analytical method.

Compare configuration conditions by actual task
Condition to compareInformation requiredSelection affected
Protection objectiveParticle source, analyte and allowable backflowFirst establish whether the method permits the protection
Component parametersPore size, filter material and check-valve opening conditionsInclude added load in the pump operating point
Maintenance planningInstallation direction, replacement and cleaning conditionsAvoid progressive blockage or contamination of the first sample

03Compare product options against task requirements

The following candidate options match this page's responsibilities. Compare them against the operating conditions described, then check capacity, materials and interfaces on the product pages. Card images show representative configurations; suitability must still be confirmed at actual operating points and by application results.

Move from task conditions to product configuration
Product categoryCorresponding roleConfirm before selection
PE-Series FiltersParticle retention permitted by the methodCards show representative configurations. Select actual pore size and filter material for the target, pressure drop and recovery requirements.
G-series check valvesBackflow limitationThe card shows a representative configuration. Confirm actual opening conditions, reverse leakage and medium separately.
  • PE-Series Filters
  • G-series check valves

04Position protection according to the source and document maintenance

Position components according to contamination or backflow sources, and record replacement and cleaning conditions. After installation, confirm orientation, priming and the first sample. Include blockage trends in maintenance planning.

Action records should link commands, valve or interface states and actual received results. When raw material, container or sample identity changes, reconfirm shared paths and conditions permitting the next step. Completion of drive movement alone does not replace task acceptance.

05Test pressure drop, backflow and analytes

Measure pressure drop, leakage or backflow, analyte loss, blanks and recovery after replacement. If the added protective load moves operation outside acceptable conditions, adjust the path and drive rather than ignoring the extra resistance.

Link operating-point tests to interpretable results
Validation scenarioMeasurement or observationInterpretation and action
Mechanical protectionMeasure pressure drop, backflow or leakageCover both protective action and normal flow
Analytical sampleMeasure analyte loss, blanks and recovery after replacementEffective filtration does not establish complete analyte retention

Record test medium, configuration, installed path and operating sequence; present routine operating points and boundary results separately. Calculations support design discussions, component documentation checks models, and installed measurements confirm the actual task.

06After an abnormality or maintenance, first identify which segment needs recovery

Clogging, adsorption, reverse leakage or excessive opening load can alter fluidics and method results. Check the affected step's actual results before resuming.

Retain current source, container identity, completed actions, valve or interface state and actual receiving information. Use the table to locate the stage that changed before deciding to retry, continue or restart. After recovery, recheck the relevant operating point before returning an uncertain action to the automatic cycle.

  • Bring the fluidic diagram and operating-point records to discuss configuration

07Continue with tasks on the same branch

  • Tubing, Fittings, Filters and Check Valves: Assess the Complete Flow Path
  • Tubing Materials and Dimensions: Volume, Resistance and Maintenance Layout
  • Low-Hold-Up Connections: Matching Threads, Tube Ends and Port Geometry

References

View references (2)
  1. [1]IDEX Health & Science: Fluidics FAQs — Dead Volume and Fittings
  2. [2]FOREACH: Fittings and Tubing Catalog

Related guides

  • Pressure and blockage monitoring: establish action baselines before fault thresholds
  • Gas-liquid waste aspiration: direct pumping or headspace vacuum
  • Low-Hold-Up Connections: Matching Threads, Tube Ends and Port Geometry

Related Products

PE-Series Filters; representative configuration. Select capacity, materials and interfaces for the application.

PE-Series Filters

G-series check valves; representative configuration. Select capacity, materials and interfaces for the application.

G-series check valves

Related Technical Articles

Miniature Diaphragm Pump Backflow After Stopping: Check Valves and Anti-Siphon Design

Diagnose reverse flow after a miniature diaphragm pump stops, distinguish it from forward siphoning and residual dripping, and evaluate check-valve behavior, valve sequencing and static sealing in the complete circuit.

2026-09-07View Details

What is flow resistance calculation for? Using the Foreach calculator in fluid-path design and selection

The free flow rate of the pump meets the requirements. After installing the pipelines, valves and joints, the liquid supply at the end is insufficient. Flow resistance calculations can break the problem down into comparable data: what pressure differential is required at the target flow rate, where does the pressure loss mainly occur, and what happens if the pipe diameter, pipe length, component Cv or medium viscosity is changed. This article combines the actual function of the Foreach flow resistance calculator with a recalculable fluid circuit case to explain how to use the calculation results for design, selection, troubleshooting and prototype verification.

2026-09-11View Details

Discuss your filter and check-valve protection fluidics

Provide a fluidic diagram, actual liquids and pressures in each operating state. Specify particle source, analyte and allowable backflow; pore size, filter material and check-valve opening conditions; and installation direction, replacement and cleaning conditions. Compare configurations against this guide's validation objectives.