Locate the source, drive and receiving position for this task
Define dose volume, dose count, first and last doses, retained volume and recovery after refill. Follow the diagram to confirm liquid contact, pressure and destination at each position before assigning drive and connection roles.
Compare displacement options against dose and cycle time
Estimate rotation and speed ranges from dose and cycle time, comparing minimum and maximum task volumes. Actual backpressure, viscosity and nozzle action may change output. Include startup and stopping states.
| Condition to compare | Information required | Selection affected |
|---|---|---|
| Individual dose and cycle time | Actual dose, pauses and container speed | Compare displacement and start-stop programs |
| Receiving outlet | Nozzle wetting, retained droplets and post-stop discharge | Set vessel-change and stop actions |
| Continuous-operation state | Changes in source level, temperature and backpressure | Check the operating range of the same configuration |
Compare product options against task requirements
The following candidate options match this page's responsibilities. Compare them against the operating conditions described, then check capacity, materials and interfaces on the product pages. Card images show representative configurations; suitability must still be confirmed at actual operating points and by application results.
| Product category | Corresponding role | Confirm before selection |
|---|---|---|
| RPL Valveless Metering Pump | Suitable repeated or continuous metering | Compare displacement, speed, actual backpressure and starts/stops; check two-stream ratios separately |
Coordinate nozzle state with vessel timing
After priming, dispense to the vessel cycle, recording nozzle droplets, remaining liquid and bottle-change timing. Check the first dose after extended standby separately. After interruption, retain records of vessels already filled.
Action records should link commands, valve or interface states and actual received results. When raw material, container or sample identity changes, reconfirm shared paths and conditions permitting the next step. Completion of drive movement alone does not replace task acceptance.
Measure startup doses and successive doses separately
Measure individual doses, first/last doses and blanks after high/low concentration changes using actual reagents. Stable continuous output does not directly establish stable start/stop dispensing.
| Validation scenario | Measurement or observation | Interpretation and action |
|---|---|---|
| Startup dose | Collect the first and subsequent doses separately after idle periods | Check nozzle and liquid-column recovery first |
| Extended dispensing runs | Measure different doses within each batch | Separate cumulative bias from individual-dose variation |
Record test medium, configuration, installed path and operating sequence; present routine operating points and boundary results separately. Calculations support design discussions, component documentation checks models, and installed measurements confirm the actual task.
After an abnormality or maintenance, first identify which segment needs recovery
Empty sources, bubbles, nozzle droplets and backpressure changes can affect first or last doses. Stop affected steps, validate recovery after priming and avoid duplicate dispensing.
Retain current source, container identity, completed actions, valve or interface state and actual receiving information. Use the table to locate the stage that changed before deciding to retry, continue or restart. After recovery, recheck the relevant operating point before returning an uncertain action to the automatic cycle.
Continue with tasks on the same branch
- Valveless metering pumps: operating conditions for repeated dispensing and continuous supply
- Valveless-pump titration: when to use bulk addition and when fine metering is needed
- Carrier and Flow-Cell Supply: Installed Flow, Bubbles and Baseline
- Dual-Stream Proportioning: Cumulative Ratio and Instantaneous Mixing