Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/Analytical instruments/Spectroscopy and elemental analysis
< Back

Liquid-Sample Elemental Analysis: Separate Standards, Dilution and Introduction

Application guide

Elemental-analysis fluidics may prepare standards, dilute overrange samples, rinse between samples or introduce samples. Metering targets and inlet conditions differ. Determine whether liquid enters a mixing vessel or goes directly to the instrument's introduction interface before selecting pumps and valves.

Instruments and methods covered

This guide covers liquid-sample elemental analysis and auxiliary fluidics. Solid, gas and other spectroscopic architectures require separate assessment of their introduction devices. Flow or capacity alone does not qualify these candidates as replacements for the original sample-introduction system.

Application guides

Spectroscopy and elemental analysis application guide

Spectroscopy and elemental-analysis fluidics overview
Syringe pumps in spectroscopy and elemental analysis
Discrete liquid-sample transfer and introduction conditionsStandards and dilution
Valveless Metering Pumps for Spectroscopy and Elemental Analysis
Multiport rotary valves in spectroscopy and elemental analysis
Multiple-source selection
Sampling needles in spectroscopy and elemental analysis
Bubble and Pressure Monitoring for Spectroscopy and Elemental Analysis
Tubing and fittings in spectroscopy and elemental analysis

ON THIS PAGE

Locate the task in the fluid path firstCommon product roles and application guidesOperating requirements differ by stepSample transfer: identify the receiving end firstStandards and dilution: include concentration in acceptance criteriaMultiple-source selection: calculate delay and displacementExplore configurations through product application guidesValidate the complete cycle through application resultsCommon selection and recovery questionsReferencesRelated guides

01Locate the task in the fluid path first

Samples and standards→Dilution or internal-standard preparation→Switching and rinsing→Instrument sample introduction→Detection and blank recovery
Illustrative task sequence. Confirm sampling, washing, waste and monitoring branches separately against the instrument drawings.

Arrows indicate task relationships, not that every component is connected in series in one tube. Check valve positions, pump wetted boundaries and operating states for each branch.

Specify acid and salt matrices, target concentrations and particle conditions, preserving actual sample-to-internal-standard volume relationships. Coordinate post-switching delay, shared-path residue and introduction conditions with the instrument method. Auxiliary pump output alone does not establish stable detection.

02Common product roles and application guides

Explore product applications by their role in the instrument
Product categoryRole in this instrument
Syringe pumpStandard preparation, dilution and internal-standard preparation
Valveless metering pumpsSuitable continuous auxiliary-liquid supply
Multiport rotary valveSelection of samples, standards, blanks and wash liquids
Sampling ProbesLiquid-sample source interface
Bubble and pressure detectionAuxiliary-path supply and pressure state
Tubing and fittingsConnections between matrix liquids and instrument interfaces
  • Syringe pumps in spectroscopy and elemental analysis
  • Valveless Metering Pumps for Spectroscopy and Elemental Analysis
  • Multiport rotary valves in spectroscopy and elemental analysis
  • Sampling needles in spectroscopy and elemental analysis
  • Bubble and Pressure Monitoring for Spectroscopy and Elemental Analysis
  • Tubing and fittings in spectroscopy and elemental analysis

03Operating requirements differ by step

Define fluid-path responsibilities before examining individual operations
StageTask categoryOperating conditions to define
Sample transfer: identify the receiving end firstSampling conditionsActual sample volume, remaining source volume, probe-tip height, aspiration/dispensing speed and system liquid
Standards and dilution: include concentration in acceptance criteriaProportion-based preparationActual volumes of original sample, diluent or internal standard; concentrations at each stage; mixing and resampling conditions
Multiple-source selection: calculate delay and displacementValve-path requirementsAll ports, valve states, pressures in each state, shared volume and flushing volume

04Sample transfer: identify the receiving end first

Delivery to a dilution vessel primarily requires volume control; delivery to an introduction interface also requires system-specific continuous flow, pulsation and pressure checks. Evaluate materials and sample-change recovery using the actual matrix.

  • Discrete liquid-sample transfer and introduction conditions

05Standards and dilution: include concentration in acceptance criteria

Measure delivery of stock, diluent and internal standard separately, then validate the resulting concentration or method response after mixing. Shared-section residue can produce an incorrect composition despite correct commanded volumes.

  • Standards and dilution

06Multiple-source selection: calculate delay and displacement

Create a port map before switching samples, standards, blanks and wash liquids. Validate blanks and stabilization time with the matrix combinations most difficult to clear, then determine displacement volume.

  • Multiple-source selection

07Explore configurations through product application guides

  • Syringe pumps in spectroscopy and elemental analysis
  • Valveless Metering Pumps for Spectroscopy and Elemental Analysis
  • Multiport rotary valves in spectroscopy and elemental analysis
  • Sampling needles in spectroscopy and elemental analysis
  • Bubble and Pressure Monitoring for Spectroscopy and Elemental Analysis
  • Tubing and fittings in spectroscopy and elemental analysis

08Validate the complete cycle through application results

Check concentration ratios, mixing, blank recovery and method results. Evaluate preparation and sample supply separately.

Plan operating-point validation and fault investigation for each step
Step to validateMeasurements or observationsPriority troubleshooting checks
Sample transfer: identify the receiving end firstMeasure actual delivery at the receiver, covering routine small volumes and maximum task volume, first dose after idle, sample changes and high/low concentration sequencesAir aspiration, clots/particles, probe contact with a wall or leakage alter actual delivery
Standards and dilution: include concentration in acceptance criteriaMeasure each branch's delivery, received ratio and mixing uniformity separately, then check assay responseSmall-volume errors, shared residual liquid or inadequate mixing may alter the actual ratio
Multiple-source selection: calculate delay and displacementCheck unintended connections, leakage, backflow and post-switching residue in each state; measure flushing volume and arrival delay through the complete pathMisalignment, shared-segment residue or abnormal switching pressure may affect the next task

A sample-preparation pump is not automatically suitable for nebulizer supply. Confirm the introduction system for the specific instrument.

09Common selection and recovery questions

Can standard preparation and instrument introduction use the same volume criterion for selection?

For preparation, verify individual liquid volumes and mixed composition. The introduction interface must separately meet actual flow, pressure and response requirements. Define the receiver before comparing drives.

Why is a delay still needed after the selector valve switches?

Check shared volume, actual flow and new-sample arrival. Validate displacement with method blanks and stabilization time; a valve reaching position does not establish that composition has updated.

References

View references (3)
  1. [1]FOREACH syringe pumps: configurations and specifications
  2. [2]FOREACH MRV3 rotary valves: configurations and specifications
  3. [3]FOREACH RPL / DRPL valveless metering pumps: configurations and specifications

Related guides

  • Syringe-pump standard preparation and dilution
  • Continuous carrier-liquid and flow-cell supply

Related Products

HLD rotary-valve syringe pump; representative configuration. Select capacity, materials and interfaces for the application.

HLD rotary-valve syringe pump

MRV3 multiport rotary valve; representative configuration. Select capacity, materials and interfaces for the application.

MRV3 multiport rotary valve

RPL Valveless Metering Pump; representative configuration. Select capacity, materials and interfaces for the application.

RPL Valveless Metering Pump

Related Technical Articles

What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

A programmable syringe pump uses a motor to control syringe plunger displacement and speed, executing aspiration, dispensing, pauses and repeated actions through commands. With a routing valve, it can connect reagent reservoirs, reaction vessels and wash paths into an automated dispensing process. Instrument development therefore needs to define volume, fluid connections and timing together.

2026-09-28View Details

Choosing Rotary Valve Wetted Materials: Head, Rotor and Seals

Identify every component in contact with liquid before comparing PEEK, PCTFE and other candidates. Evaluate chemical compatibility, dynamic sealing and analytical results together.

2026-09-30View Details

Valveless Metering Pump Selection Guide: FOREACH Models and Applications

To select a valveless metering pump, first define the volume per dose and the time available for dispensing. Then check displacement per revolution, drive speed, wetted materials, backpressure and connections, and verify delivery in the actual fluid path. Neither flow rate alone nor the label ‘metering pump’ is enough to select a model.

2026-09-16View Details

Discuss fluidics for your liquid-sample elemental analysis

Provide a fluid diagram, the actual liquids and pressures in each state. Specify volume ratios and mixing, shared sections and matrix compatibility, and introduction conditions and signal stability. Compare suitable configurations against the validation objectives described here.

Contact an engineer