Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control Series
Applications
IVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Fluidic Components for
Analytical Instruments

Pumps, valves, probes, fittings, tubing and sensing components for precise sample preparation, reagent handling and instrument fluidics.

Key Fluidic Tasks

  • Sample aspiration and metering
  • Reagent delivery and path switching
  • Rinsing, drainage and waste handling
  • Pressure, liquid-level and bubble monitoring
Home/Applications/Analytical Instruments
  • Chromatography / Autosampler
  • Spectroscopy / Elemental Analysis
  • Water Quality / Environmental Monitoring
  • Sample Preparation Systems
  • Laboratory Analyzer System Integration

Piston Pumps for Sample Aspiration and Transfer

Application guide · Clinical chemistry sample handling

FOREACH piston pumps provide the metering motion for sample aspiration and transfer in clinical chemistry analyzers. Working with a sampling probe, valves and a system-fluid channel, they deliver serum, assay-approved plasma or urine to the reaction vessel. For small samples, the configuration must address working stroke, the liquid segment at the probe and washing: the required volume must arrive without unintended dilution or unacceptable carryover from the previous sample.

Consider a fixed-probe channel that normally delivers 5 μL and also needs a 20 μL dose. The following engineering example compares 50 μL and 100 μL capacities, system-fluid displacement, slow aspiration response, the first dose after idle and sample-change residue. The 5 μL and 20 μL points are design examples; the assay determines the instrument's actual volumes. EA offers capacity, material and drive options; SM and TM can be compared where space is constrained, while EAS addresses a separate need for easier pump-chamber deaeration.

Application guides

Analytical instruments

FOREACH in analytical instruments
Piston pumps in analytical instruments
Sample aspiration and transferReagent and standard dispensingDilution, spiking and proportioningIncremental titrant dosing
Syringe pumps in analytical instruments
Automated sampling and injectionMultichannel reagent distributionAutomated dilution and standardsSequential injection and flow analysis
Pipetting pumps in analytical instruments
Disposable-tip sample transferTip-based dilution and distributionSample-preparation workflows
Valveless metering pumps in analytical instruments
Repeated reagent dispensingTitrant meteringCarrier and flow-cell supplyProportional liquid preparation
Diaphragm pumps in analytical instruments
Wash-liquid delivery and primingWater and working-liquid replenishmentThermal and auxiliary circulationGas-liquid waste aspiration
Fluidic valves in analytical instruments
Multi-source flow selectionSample-loop injection and high-pressure switchingDetector diversion, bypass and wasteTimed shutoff and isolation
Probes and mixing in analytical instruments
Sampling and liquid deliveryProbe-path washingReaction-vessel and microplate washingReaction mixing
Fluid-state monitoring in analytical instruments
Bubble and liquid-state detectionPressure and blockage monitoring
Tubing and connections in analytical instruments
Tubing material and dimensionsFittings and assemblyFiltration and backflow protection

ON THIS PAGE

Define the sample and its contact path firstCompare 5 μL and 20 μL doses with the complete stroke budgetKeep the sample at the front end and deliver the aliquot to the vesselSeparate slow aspiration, clots, analytical interference and bubblesVerify volume, sample integrity and recovery at the receiving positionReferencesRelated guides

01Define the sample and its contact path first

A sample channel must obtain a defined aliquot from a limited specimen, preserve its composition and deliver it to the reaction position. First decide where the sample will remain. With system-fluid displacement, the sample can stay mainly in the probe and front-end tubing while the pump contains system fluid. If sample passes through the pump, the pump, valves and tubing all become sample-contact surfaces that require cleaning. This decision comes before selecting the pump-head material or wash volume.

Specimen and aspiration conditions are assay-specific
SampleChannel design considerationsInputs needed
Separated serumWhere the sample segment resides; small-volume delivery; internal and external probe washing; carryover.Minimum and routine doses, available specimen volume, and abnormal protein content or visible microclots.
Assay-approved plasma, such as heparin plasma for applicable testsAnticoagulant, fibrin and specimen condition as well as volume.The anticoagulant and target assay. Acceptance for one assay does not establish suitability for every chemistry test.
Urine for the specified assayDirect sampling or predilution; sediment, mixing and aspiration position.The method's preparation, diluent, dilution ratio and aliquot volume.

For example, the Roche GLUC3 method accepts serum, specified plasma types and urine under its respective method settings. FOREACH configures the metering channel around these defined conditions. When system fluid drives the sample, list the pump-contact liquid separately from the specimen contacting the probe.[1]

Available specimen volume is not the same as the delivered dose. A 5 μL reaction dose does not mean that a tube containing only 5 μL can be sampled reliably. Probe immersion, inaccessible liquid at the container bottom, aspiration reserve and repeat tests all consume the specimen budget. For limited specimens, reduce unusable front-end volume and unnecessary aspiration loss first. This page concerns serum, plasma and urine prepared as required by the assay; cellular whole blood and disposable-tip pipetting require their own fluid-path design.

02Compare 5 μL and 20 μL doses with the complete stroke budget

The assay determines the working dose. Roche GLUC3 on cobas c 303, c 503 and c 703 uses a normal sample volume of 1.5 μL. That method point illustrates an even shorter stroke; it is not a FOREACH small-volume performance claim. The 5 μL and 20 μL points below are an independent capacity-selection example.[1]

Nominal conversion for EA configurations with 2,000 full-stroke steps[2]
Delivered-volume targetEA 50 μL: stroke / stepsEA 100 μL: stroke / stepsSelection implication
1.5 μL3% / 601.5% / 30Evaluate reversal, valve motion and probe-tip delivery at this short working stroke.
5 μL10% / 2005% / 100The smaller capacity uses more stroke; check that auxiliary aspiration segments still fit.
20 μL40% / 80020% / 400Compare 50 μL first if both doses are primary tasks, then check the maximum uninterrupted aspiration budget.

Stroke percentages and step counts are calculated command relationships, not measured liquid-delivery accuracy.

The nominal full-step volumes are 0.025 μL for the 50 μL configuration and 0.05 μL for the 100 μL configuration. More full steps for the same dose permit finer command increments, but transmission behavior, reversal, valve switching and tip retention still affect delivery. Accuracy does not improve in direct proportion to the number of steps.[2]

If a project sets a mean-volume error target of ±1% at 5 μL, the absolute allowance is ±0.05 μL—one nominal full step of the 100 μL / 2,000-step configuration. This comparison calls for both mechanical and liquid-delivery evaluation; the availability of motor microstepping alone does not establish compliance. The overview explains why full-stroke specifications and the datasheet's 2%-stroke displacement tests cannot replace validation of this delivered dose.

Include auxiliary movements in the same uninterrupted stroke. For example, 20 μL of useful sample, a nominal 5 μL air gap and 5 μL of auxiliary displacement total 30 μL, leaving 20 μL of nominal capacity in a 50 μL pump. Adding another 30 μL segment in that stroke raises the budget to 60 μL; compare a 100 μL configuration or change the sequence. These auxiliary volumes are calculation assumptions, not standard sample-program settings. Check gas compression and segment position under the actual pressure and sequence.

Limited specimen volume calls for reduced front-end waste. Large auxiliary segments call for a stroke-budget check. Restricted installation space calls for comparison with SM or TM. These are different constraints and do not all lead to a larger pump.

  • Piston-pump capacity and performance selection

03Keep the sample at the front end and deliver the aliquot to the vessel

For patient samples that change from tube to tube, a FOREACH piston pump can move system fluid while the sample remains mainly in the probe and front-end tubing. This limits the extent of sample contact with the pump, but does not remove interface mixing, tip retention or wash residue. Direct aspiration into the pump is also possible in a suitable dedicated fluid path, provided every wetted surface can be cleaned and sample composition is preserved.

Turn the sampling cycle into defined liquid-handling actions
ActionFluid-path and program arrangementRequired outcome
PrimeFill the pump, valves and drive tubing with the specified system fluid; send recovery liquid to waste.Establish a stable liquid column before metering. Priming volume is not sample volume.
AspirateCoordinate liquid level, aspiration speed and settling time; define any method-approved air gap separately.Acquire the complete sample segment without aspirating air or excessive system-fluid mixing.
DeliverCoordinate valve position, dispensing, contact or non-contact transfer and subsequent retraction.Deliver the required sample to the reaction vessel, not merely complete the motor command.
Change samplesWash the probe internally and externally and clean the actual shared path.Control the previous sample's contribution to the next sample.

Research on liquid-filled automated handling identifies liquid segments, air gaps and motion speed as variables affecting sample integrity. In a simple 5 μL delivery example, replacing 0.1 μL of sample with analyte-free system fluid lowers concentration by 2%, assuming additive volumes and no analyte loss. Measuring a total volume of 5 μL therefore does not, by itself, confirm the effective sample fraction.[3]

Fix the tip-delivery arrangement early. Contact dispensing uses the vessel wall or existing liquid to assist droplet transfer, so contact position and subsequent cleaning matter. Non-contact dispensing requires the drop to detach completely; dispensing speed, outlet geometry and liquid surface properties work together. If the final drop remains on the tip, extra motor steps may affect the next delivery rather than repair the current one. Tune retraction as a separate action after the effective dose has transferred.

A deliberate air gap is different from an uncontrolled bubble in the pump. The former is a defined part of the segment program; the latter can consume displacement as pressure changes. When composition is wrong, inspect interfaces and delivery order. When volume fluctuates under an unchanged program, investigate air aspiration, leaks and trapped gas. Metering motion, valves and probe handling must be tuned against the same receiving position; motor completion and valve confirmation are not substitutes for the final liquid result.

04Separate slow aspiration, clots, analytical interference and bubbles

Distinguish liquid that has not reached its intended position in time from a specimen that is unsuitable for the assay. Fluid-path and motion changes may improve the former; the latter requires the method's specimen-handling rules. Treating every abnormal sample as a viscosity problem leads to the wrong compensation.

Match the response to the actual cause[1][4][5]
Observed conditionConfiguration or instrument responseDecision boundary
Serum or plasma with abnormal immunoglobulin content and slow aspiration responseEvaluate inlet resistance at the actual temperature; adjust aspiration speed, acceleration and settling independently.Confirm under-aspiration. Protein precipitation or assay interference is a separate issue that slower pumping does not resolve.
Fibrin strands or microclotsUse appropriate instrument clot detection and exception handling; apply the required specimen preparation before resampling.Do not use increased drive force to pull through a blockage. The pump does not replace specimen preparation.
Lipemic, hemolyzed or icteric specimensCoordinate with sample-integrity checks and assay-specific interference rules.Appearance or optical interference is not equivalent to high viscosity; changing dose or pump speed does not establish an acceptable assay result.
High-concentration sample followed by a low-concentration sampleReduce the shared sample path and apply internal, external and, where needed, targeted probe washes.Measure the effect on the low sample, not only the volume of a water blank.
Gas trapped in the system-fluid pump after a liquid changeCompare EAS chamber geometry together with orientation, inlet sealing and a priming sequence.Easier chamber deaeration does not correct an empty inlet or continuing upstream gas release.

For confirmed slow aspiration without a blockage, first check inlet probe and tubing length and bore. Reduce aspiration speed, soften acceleration and allow the liquid column to settle while the probe remains in its specified position. Tune dispensing separately instead of slowing the entire cycle proportionally. These changes consume time: if volume recovers but throughput fails, improve the inlet path or redistribute the cycle time. Tecan's pump-selection guidance likewise relates liquid behavior to aspiration, dispensing, settling and throughput.[6]

A stable low mean under repeatable conditions may justify working-point calibration after residue and dilution have been excluded. An abnormal first dose followed by normal doses points to idle recovery or priming. Occasional large deficits point first to clots, air aspiration or bubbles. A fixed volume offset is not an effective correction for a changing fault. EAS supports chamber deaeration; it does not provide specimen-clot or low-level detection.

Carryover can be translated into an assay error budget. If a 3 mmol/L glucose segment has 0.1% of its volume replaced by a preceding 30 mmol/L sample, an ideal mixing calculation gives an increase of (30 − 3) × 0.001 = 0.027 mmol/L, or 0.9% of the low value. This is a worked error-budget example, not a universal carryover limit; the assay defines the acceptable effect.

  • Aspiration and dispensing motion profiles
  • Priming and bubble-related dispensing error

05Verify volume, sample integrity and recovery at the receiving position

Connect the metering configuration to the fixed-probe channel
AreaConditions to defineWhen to reconsider the arrangement
Pump and front-end wetted pathConfigure the pump for the actual system fluid and cleaners passing through it; configure the probe for specimen, cleaners and aspiration position.If sample enters the pump, include its chamber and valves in compatibility, adsorption and washing evaluation.
Tip and valve pathDefine tip position, shared-path volume, internal and external washing together with the useful sample segment.If high-to-low carryover exceeds the assay allowance, locate residue before comparing dedicated paths or more effective washing.
Control and recoverySpecify homing, optional position feedback, valve states and priming to waste separately.Air aspiration and blocked-probe detection need relevant sensing and exception logic; an encoder alone cannot identify them.

EA offers an optional ISC1000 driver with RS-232, RS-485 or CAN interface configurations. Initial-position sensing and optional motor closed-loop optical encoder feedback are distinct options. FOREACH can match the metering drive, interfaces and valve components to the application; probe motion and assay coordination remain instrument-level functions.[2]

For an illustrative acceptance target of ±1% mean-volume error and CV ≤1% at 5 μL, the acceptable mean is 4.95–5.05 μL and standard deviation is no more than 1% of the measured mean. A ±1% mean-volume target at 20 μL corresponds to 19.8–20.2 μL. These are project acceptance examples, not new small-dose product specifications. Define accuracy, repeatability and CV consistently with the selected test procedure.

Assess continuous sampling, the first dose after the longest planned idle period, the first dose after sample-change washing and sampling at low liquid level separately. Volume measurement establishes how much arrived; an appropriate tracer or analytical method checks unintended system-fluid dilution; a high-to-low sequence measures carryover. Verification of a liquid handler must cover the actual task and sample integrity as well as volume.[7][3]

At 5 μL, a ±1% allowance is only ±0.05 μL. Record the specimen or surrogate, temperature, density, replicate count and measurement uncertainty, and control evaporation. Balance display resolution alone does not establish that the measurement method is adequate.

For configuration discussions, provide a concrete brief: fixed-probe serum channel; 5 μL routine and 20 μL maximum dose; identified system fluid and cleaners; auxiliary aspiration segments; complete cycle time; and separate first-dose and steady-state bias/CV targets. This narrows the choice to a useful capacity, wetted assembly and drive configuration.

References

View references (7)
  1. [1]Roche — Glucose HK Gen.3, cobas c 303 / c 503 / c 703, V7.0 (2024)
  2. [2]FOREACH — EA Piston Pump Datasheet, basic capacities and test conditions
  3. [3]Ouyang et al. — Strategies to Maintain Sample Integrity Using a Liquid-Filled Automated Liquid-Handling System with Fixed Pipetting Tips (2008)
  4. [4]Hortin et al. — Problematic Proteins: A Patient with a High Paraprotein Concentration (2024)
  5. [5]FOREACH — EAS Easy-to-Degas Piston Pump Datasheet
  6. [6]Tecan — Liquid handling pump selection: a guide for lab automation engineers
  7. [7]Albert and Bradshaw — Importance of Integrating a Volume Verification Method for Liquid Handlers: Applications in Learning Performance Behavior (2007)

Related guides

  • Piston-pump capacity and configuration
  • Predilution and resampling
  • Reagent dispensing and first-dose recovery

Related Products

EA-500-PMMA

EA-500-PMMA

EA-500-PEEK

EA-500-PEEK

SM-100-PMMA

SM-100-PMMA

TM-100-PMMA

TM-100-PMMA

Related Technical Articles

Why Do Air Bubbles Delay or Bias Piston Pump Dispensing?

A completed piston stroke does not prove that the intended liquid volume reached the vessel. Bubble compression temporarily stores displacement; expansion can then cause a trailing dispense or affect the next cycle. Locate where bubbles first appear and compare synchronized pressure and individual dispense masses before and after priming.

2026-09-10View Details

How Should Piston Pump Aspiration Speed and Dwell Be Set for Viscous Reagents?

A stable aspiration window depends on resistance, filling time and pressure. Test aspiration speed and the subsequent dwell independently, then select a faster combination that satisfies bias, repeatability, bubbles and cycle-time requirements. Water settings do not automatically transfer to the reagent, and completed motion does not prove fluid equilibration.

2026-09-10View Details

How Can You Isolate Piston Pump Dispensing Drift Over Time?

A volume trend shows that system conditions are changing; it does not identify pump wear. Save every dispense with temperature, level, reversal and priming events, then isolate causes using controlled priming, fixed level and pump/valve/circuit swaps. Review mean bias separately from CV before concluding that mechanical parts or seals have degraded.

2026-09-10View Details

Choosing a 100, 250 or 500 μL Piston Pump for Clinical Chemistry Dosing

Compare dose volume, stroke utilization, dispensing count and refill timing, then assess Foreach EA, SM and TM configurations using the actual clinical chemistry fluid path.

2026-09-08View Details

How to Select a Piston Pump Head Material

Compare acrylic (PMMA), PCTG, PPS, PVDF, PP, PTFE, PEEK, acetal (POM) and polysulfone (PSU) pump-head options, and learn how to validate the complete wetted path.

2026-09-06View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Precision Piston Pump? Working Principle, Benefits and Applications

Learn how precision piston pumps work, where they are used and how to select capacity, performance, materials, valves and controls for an OEM fluidic system.

2026-06-19View Details

Acceleration and Deceleration Curves for Piston-Pump Drives

Use stepper-motor torque-speed behavior to select linear, S-curve or parabolic motion profiles and prevent missed steps in piston-pump aspiration and dispensing.

2026-09-06View Details

Precision Piston Pump Backlash Compensation Guide

Understand reversal backlash, measurement methods and compensation timing, and distinguish mechanical lost motion from actual aspirated or dispensed volume.

2026-09-06View Details

Precision Piston Pump Stepper Motor Calculation and Selection

Calculate pulse frequency, microstepping, transmission ratio and load torque, then verify speed and dynamic torque with a 400 mm reciprocating-platform example.

2026-09-06View Details

Need help with an analytical instrument fluid path?

Share the medium, flow range, pressure, connection and control requirements with our engineering team.