Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control Series
Applications
IVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Fluidic Components for
Analytical Instruments

Pumps, valves, probes, fittings, tubing and sensing components for precise sample preparation, reagent handling and instrument fluidics.

Key Fluidic Tasks

  • Sample aspiration and metering
  • Reagent delivery and path switching
  • Rinsing, drainage and waste handling
  • Pressure, liquid-level and bubble monitoring
Home/Applications/Analytical Instruments
  • Chromatography / Autosampler
  • Spectroscopy / Elemental Analysis
  • Water Quality / Environmental Monitoring
  • Sample Preparation Systems
  • Laboratory Analyzer System Integration

Piston Pumps for Incremental Titrant Dosing

Application guide · Automated titration

A FOREACH EA piston pump can form a programmable dosing channel with a titrant reservoir, switching valve and titration tip. In aqueous acid–base potentiometric titration, EA provides metered aspiration and dispensing with a configurable drive. The instrument uses the electrode signal to coordinate additions, mixing, readings and stopping; pump motion must follow those states.

Consider standardized 0.1 mol/L aqueous NaOH or HCl, an expected consumption near 2 mL and a design aim to compare increments around 4 μL near the endpoint. These are inputs to the worked example, not universal acid–base titration settings. The method determines reagent preparation, concentration and endpoint criteria.

Application guides

Analytical instruments

FOREACH in analytical instruments
Piston pumps in analytical instruments
Sample aspiration and transferReagent and standard dispensingDilution, spiking and proportioningIncremental titrant dosing
Syringe pumps in analytical instruments
Automated sampling and injectionMultichannel reagent distributionAutomated dilution and standardsSequential injection and flow analysis
Pipetting pumps in analytical instruments
Disposable-tip sample transferTip-based dilution and distributionSample-preparation workflows
Valveless metering pumps in analytical instruments
Repeated reagent dispensingTitrant meteringCarrier and flow-cell supplyProportional liquid preparation
Diaphragm pumps in analytical instruments
Wash-liquid delivery and primingWater and working-liquid replenishmentThermal and auxiliary circulationGas-liquid waste aspiration
Fluidic valves in analytical instruments
Multi-source flow selectionSample-loop injection and high-pressure switchingDetector diversion, bypass and wasteTimed shutoff and isolation
Probes and mixing in analytical instruments
Sampling and liquid deliveryProbe-path washingReaction-vessel and microplate washingReaction mixing
Fluid-state monitoring in analytical instruments
Bubble and liquid-state detectionPressure and blockage monitoring
Tubing and connections in analytical instruments
Tubing material and dimensionsFittings and assemblyFiltration and backflow protection

ON THIS PAGE

Select capacity from both total consumption and the endpoint incrementA 4 μL increment is not a titration-accuracy specificationCoordinate variable increments, constant increments and endpoint controlDefine wetted materials and maintenance for the actual aqueous titrantEnsure each increment reaches the reaction before taking the readingChoose one pump or coarse/fine channels from consumption and cycle requirementsReferencesRelated guides

01Select capacity from both total consumption and the endpoint increment

A larger capacity reduces refills; a smaller capacity gives the same increment a longer stroke. An expected 2 mL total does not, by itself, justify a large pump, and a microliter endpoint increment does not automatically require the smallest one. Compare total consumption, minimum increment and drive configuration together.

EA standard-step comparison: 2 mL total and 4 μL increments are design assumptions[1]
EA configurationNominal relationship for 4 μLCapacity budget for 2 mLSelection condition
100 μL / 2,000 steps0.05 μL per full step; 80 steps; 4% stroke.Ideally 20 full-stroke dispensing segments: at least 19 refills after the initial fill.Fine command increments, provided the method can accommodate frequent refill and recovery.
500 μL / 2,000 steps0.25 μL per full step; 16 steps; 0.8% stroke.Ideally four full-stroke dispensing segments: at least three refills after the initial fill.Fewer refills, but the 4 μL partial-stroke action and tip delivery still need separate evaluation.
2.5 mL / 2,000 steps1.25 μL per full step; 4 μL equals 3.2 full steps.2 mL uses 80% of stroke, leaving 0.5 mL nominal capacity.Integer full steps cannot represent exactly 4 μL. Compare another drive configuration or a method-permitted increment verified at the tip.

These ideal capacity budgets exclude predispense, reserve and refill recovery. Deduct any same-stroke auxiliary volumes from usable capacity.

For an approximately 2 mL determination, the 2.5 mL configuration leaves a nominal 0.5 mL margin, which must also accommodate any predispense and reserve using that stroke. Other resolutions can be considered, but a finer command still needs to be evaluated against mechanical response and the delivered liquid increment.[1]

Metrohm's guidance on transferring manual titration to autotitration recommends an expected endpoint volume within 10–90% of buret capacity. It illustrates why total consumption and refill planning matter alongside fine increments. That recommendation belongs to its titration-method development context; it is not a guaranteed accuracy range for every FOREACH pump.[2]

The 2.5 mL row uses the EA basic-capacity table's 2,000-step, 1.42 mm screw-lead configuration. EA-2500 also has a 2,236-step, 1.27 mm lead configuration, giving approximately 1.118 μL per full step. Match the ordered version, drive conversion and software volume accumulator. Do not interchange the two step counts or substitute the EAS-2500 configuration without recalculation.[1][3]

02A 4 μL increment is not a titration-accuracy specification

4 μL ÷ 2.00 mL = 0.2%, while 20 μL ÷ 2.00 mL = 1%. These ratios help judge whether an increment may be too coarse relative to expected consumption; they are not the accuracy of the titration result. Fixed-endpoint stopping, dynamic equivalence-point detection and curve evaluation use the data differently. Reaction kinetics, mixing and electrode response also affect the result.

For the 2.5 mL / 2,000-step configuration, three full steps represent 3.75 μL and four represent 5 μL. If the controller executes three steps but adds 4 μL to the recorded total, that action has a 0.25 μL command-label discrepancy. Keep the volume conversion, drive configuration and accumulated record consistent. Where variable increments are method-permitted, record each commanded increment; an average of 4 μL is not a valid substitute for every individual action's record.

Validate the selected configuration at the intended small-increment working point with its valves, tubing and tip. Full-stroke specifications—and the separate displacement-based short-stroke tests—help define the component's stated conditions, but cannot replace this liquid-delivery result.

  • Full-stroke specifications and short-stroke test boundaries

03Coordinate variable increments, constant increments and endpoint control

Metrohm distinguishes dynamic equivalence-point titration (DET), monotonic equivalence-point titration (MET) and endpoint titration (SET). DET varies the volume increment with the curve, MET uses constant increments, and SET controls toward a predefined signal endpoint. Readings can depend on waiting time or signal-drift conditions. The reaction and method determine the mode; running the pump continuously at one speed does not perform all these functions.[4]

Separate FOREACH dosing actions from instrument-level method decisions
Operating stateEA and valve actionInstrument decision and record
Prime before the runAspirate titrant and route gas-clearance and displacement liquid to waste.Confirm a stable column at the tip. Waste volume is not titration consumption.
Early additionExecute the method-permitted initial dose or larger increments; record command volumes for the actual configuration.Maintain mixing and determine when a denser set of measurement points is needed.
Near the endpointDeliver the requested small increment, then hold the defined valve and tip state.Wait for mixing and electrode response; record each addition, signal and time. Do not treat response lag as a reason to continue a large addition.
Refill and recoverIsolate the titration-vessel path, refill and perform any required gas clearance or predispense.Accumulate only the portion delivered into the vessel. Where the method permits, schedule refill windows instead of interrupting unexpectedly near the endpoint.
StopEnd useful dispensing and manage retained drops or retraction.Record the endpoint criterion and confirm that subsequent dripping does not continue the reaction.

The FOREACH drive executes displacement, speed and reversal; electrode acquisition and endpoint recognition belong to the instrument. Optional ISC1000 configurations with RS-232, RS-485 or CAN connect the dosing channel to the controller. Optional closed-loop position feedback is not a feedback loop for the titration reaction itself.[1]

04Define wetted materials and maintenance for the actual aqueous titrant

This example uses standardized 0.1 mol/L aqueous NaOH or HCl. Specify concentration, temperature, contact duration and cleaners rather than describing the service only as acid or alkali. Non-aqueous titration, organic-solvent systems and Karl Fischer reagents are different applications and must not inherit this aqueous configuration without evaluation.

EA supports different head and piston configurations; its datasheet lists a ceramic piston, UHMWPE seal and optional seal flushing. Compatibility must extend beyond the head to seals, valves, tubing and fittings under their actual exposure. For salt-containing or crystallization-prone residue, establish cleaning and shutdown management at the affected surfaces rather than masking rising friction indefinitely with more drive force.[1]

For an insufficient first dose after a bottle change, first address the inlet air segment, bottle supply and priming. If chamber gas remains difficult to clear, compare EAS. If air enters on every refill, inspect level and inlet sealing at the source. EAS improves the chamber's gas-clearance path; it does not control titrant concentration or guarantee a bubble-free complete fluid path.[5]

  • Pump-head, piston and complete wetted-material selection

05Ensure each increment reaches the reaction before taking the reading

Liquid retained on the tip after pump motion stops has not all entered the titration vessel. A drop that falls after the electrode reading separates the recorded event from the reaction event. When evaluating a 4 μL-scale increment, fix tip geometry, position and relationship to the liquid surface; observe isolated increments, continuous additions and the first dose after a pause separately.

Above-surface delivery requires timely droplet detachment. An immersed tip also requires management of back-diffusion, sample contact and cleaning. Choose the position according to the method and tip design. Excessive retraction can draw in air or make the next command first restore the liquid column. Evaluate retraction against the following dose, not only whether the tip appears drip-free.

The full cycle includes dosing, mixing, electrode response and any refill. Shorter motor travel time does not necessarily shorten the determination. If the increment has reached the vessel but the signal is still changing, follow the method's wait or exception rules instead of removing the wait indiscriminately.[4]

Gravimetric evaluation of microliter increments needs an appropriate balance, density correction and evaporation control. A cumulative weight from many doses helps assess total volume but can conceal erratic individual doses or a deficient first dose. After checking liquid delivery, use an appropriate reference material under the method to evaluate the complete titration result.

06Choose one pump or coarse/fine channels from consumption and cycle requirements

For low total consumption with fine increments and acceptable refill opportunities, begin by comparing small-capacity EA channels. For consumption approaching several milliliters where frequent refill is undesirable, compare 2.5 mL or another suitable capacity with its actual drive configuration. Covering the total volume does not mean that standard full steps can express any desired small increment. Match the drive and tip at the minimum useful increment.

If early bulk addition consumes much of the cycle while the endpoint still requires fine dosing, compare two coarse/fine channels: a larger capacity handles the method-permitted early addition and a smaller capacity handles the final region. Both require calibration, independent isolation and metering records; include the additional valves and maintenance cost. A second channel is unnecessary if one pump already meets increment and cycle requirements.

Provide titrant identity and concentration, minimum/routine/maximum consumption, minimum useful increment, method mode, allowed cycle time, whether refill interruptions are permissible, tip position and cleaning requirements. Keep records of individual increments, cumulative delivery, the first dose after pause or refill, and the full titration's endpoint performance. FOREACH can then match capacity, wetted assembly, valves and drive instead of selecting a model solely from a 2 mL total.

References

View references (5)
  1. [1]FOREACH — EA Piston Pump Datasheet, capacities, drive options and test conditions
  2. [2]Metrohm — How to transfer manual titration to autotitration (2026)
  3. [3]FOREACH — EA-2500-PEEK transmission configurations
  4. [4]Metrohm — Eco Titrator manual, §6.12: DET, MET and SET parameters
  5. [5]FOREACH — EAS Easy-to-Degas Piston Pump Datasheet

Related guides

  • Piston-pump capacity and configuration
  • Reagent dispensing and idle recovery
  • Dilution, spiking and proportioning

Related Products

EA-500-PMMA

EA-500-PMMA

EA-500-PEEK

EA-500-PEEK

SM-100-PMMA

SM-100-PMMA

TM-100-PMMA

TM-100-PMMA

Related Technical Articles

Why Do Air Bubbles Delay or Bias Piston Pump Dispensing?

A completed piston stroke does not prove that the intended liquid volume reached the vessel. Bubble compression temporarily stores displacement; expansion can then cause a trailing dispense or affect the next cycle. Locate where bubbles first appear and compare synchronized pressure and individual dispense masses before and after priming.

2026-09-10View Details

How Should Piston Pump Aspiration Speed and Dwell Be Set for Viscous Reagents?

A stable aspiration window depends on resistance, filling time and pressure. Test aspiration speed and the subsequent dwell independently, then select a faster combination that satisfies bias, repeatability, bubbles and cycle-time requirements. Water settings do not automatically transfer to the reagent, and completed motion does not prove fluid equilibration.

2026-09-10View Details

How Can You Isolate Piston Pump Dispensing Drift Over Time?

A volume trend shows that system conditions are changing; it does not identify pump wear. Save every dispense with temperature, level, reversal and priming events, then isolate causes using controlled priming, fixed level and pump/valve/circuit swaps. Review mean bias separately from CV before concluding that mechanical parts or seals have degraded.

2026-09-10View Details

Choosing a 100, 250 or 500 μL Piston Pump for Clinical Chemistry Dosing

Compare dose volume, stroke utilization, dispensing count and refill timing, then assess Foreach EA, SM and TM configurations using the actual clinical chemistry fluid path.

2026-09-08View Details

How to Select a Piston Pump Head Material

Compare acrylic (PMMA), PCTG, PPS, PVDF, PP, PTFE, PEEK, acetal (POM) and polysulfone (PSU) pump-head options, and learn how to validate the complete wetted path.

2026-09-06View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Precision Piston Pump? Working Principle, Benefits and Applications

Learn how precision piston pumps work, where they are used and how to select capacity, performance, materials, valves and controls for an OEM fluidic system.

2026-06-19View Details

Acceleration and Deceleration Curves for Piston-Pump Drives

Use stepper-motor torque-speed behavior to select linear, S-curve or parabolic motion profiles and prevent missed steps in piston-pump aspiration and dispensing.

2026-09-06View Details

Precision Piston Pump Backlash Compensation Guide

Understand reversal backlash, measurement methods and compensation timing, and distinguish mechanical lost motion from actual aspirated or dispensed volume.

2026-09-06View Details

Precision Piston Pump Stepper Motor Calculation and Selection

Calculate pulse frequency, microstepping, transmission ratio and load torque, then verify speed and dynamic torque with a 400 mm reciprocating-platform example.