What do these products do in this instrument?
Bubble and pressure detection: identify supply and transfer anomalies
Collect pressure or bubble signals in the actual branch. The controller associates them with the current source, workstation and delivery stage. Use different restart rules for partial and completed transfers.
ABD provides bubble or liquid-state signals for defined tubing; PDM5 provides pressure information at a defined location. Configure module, position and action window separately; the controller associates affected samples or tasks.
Match sensing locations to specific fluidic problems
Bubble modules report bubble or liquid state in defined tube segments; pressure modules report pressure at defined positions. Before selecting mounting points, define the phase and abnormality to observe: aspiration, delivery, source depletion or branch-load changes. Associate sensor positions with affected samples, cuvettes or stations.
Connect signals, action windows and recovery conditions
The same signal may mean different things at different valve positions or operating phases. Record normal ranges and faults with actual liquids, tubing and sequences. The controller combines action timing, confirmed delivery status, volume records and sample identity to pause, review or recover; one pressure change cannot directly identify a blockage location or specific cause. Distinguish measured, estimated and unknown volumes; completion of a command does not prove actual liquid delivery.
Translate actual branch conditions into configuration requirements
| Condition | To confirm |
|---|---|
| Installation conditions | Specify tubing dimensions, liquid, sensing location and branch to observe. |
| Timing and baseline | Record valve positions, pump actions, normal changes and fault-detection windows. |
| Abnormality handling | Associate affected containers with confirmed delivery status and volume records; define conditions for review, continuation or re-execution. Distinguish measured, estimated and unknown volumes; completion of a command does not prove actual liquid delivery. |
| Operation | Specific operating conditions |
|---|---|
| Washing and product recovery | Solid-phase location and retention target, wash cycles, aspiration height, residue, destination and downstream assay requirements |
Compare product configurations by the roles on this page
| Product | Configuration option | Conditions to verify |
|---|---|---|
| ABD bubble-detection module | Bubble or liquid-state identification in defined tubing | Validate detection and false alarms for tubing material, dimensions, medium and action window |
| PDM5 pressure-detection module | Pressure-state monitoring at a defined fluidic location | Confirm range and wetted configuration by model; validate normal baseline and abnormalities separately |
Related operations and architecture comparisons
Washing and product recovery
When several stations wash simultaneously, shared supply and waste removal can become bottlenecks. Calculate peak demand first, then assess residual liquid and recovery through the complete workflow.
Confirm product roles in actual operation
| Operation | Measurements and observations | Priority checks |
|---|---|---|
| Washing and product recovery | Evaluate solid-phase loss, post-wash residue and downstream assay background or nucleic-acid recovery together | Insufficient capture, incorrect aspiration height, uneven coverage or pooling may impair washing and recovery |
Validate working volume, timing across workstations, sample identity and the first operation after restart together.
Can dispensing be repeated immediately after bubbles or abnormal pressure are detected?
First establish whether partial delivery has occurred and which container is affected. Immediate repetition may double-dose it. Act on the actual state and recheck the first cycle after recovery.
Successful platform operation does not replace assessment of nucleic acid, cell or analytical method results.