What do these products do in this instrument?
Syringe pumps: reagent metering through fixed paths
In fixed fluidics, aspirate and dispense the reagents required by the method, coordinated with source selection and priming. Where magnetic rods transfer the solid phase, pumps serve only the replenishment or liquid-transfer branches actually present.
HLD and HMD organize aspiration/dispensing paths with rotary and solenoid valves respectively. Capacity budgets include useful volume, holding stages and replenishment points; valve states and pump actions together create each delivery.
Organize syringe metering around aspiration, holding and dispensing
Syringe drives use controlled displacement for aspiration, holding and dispensing. With direct contact, source liquid enters the metering chamber; with system-liquid drive, sample can remain at the probe tip or a designated tube segment. Define wetted coverage before capacity, cleaning and replenishment. For repeated dispensing, track current inventory, delivered volume and next replenishment time.
Map valve connections to every action
Check port connections, opening/closing sequence and power-off state separately for rotary- and solenoid-valve configurations. Changing source does not automatically remove old liquid from the chamber or shared tubing. Plan flushing for formulation and residue requirements, then evaluate the first dose after switching and dose-to-dose variation.
Translate actual branch conditions into configuration requirements
| Condition | To confirm |
|---|---|
| Stored volume and timing | Define volume per operation, consecutive dispensing count, holding time and permitted replenishment window. |
| Valve positions and destinations | Provide a source, receiver and waste port diagram with open/closed and power-off states. |
| Liquid changes and first dose | Check shared residue, priming destination and actual received volume after standby or source changes. |
| Operation | Specific operating conditions |
|---|---|
| Sample preparation | Dose per operation, total cycle volume, replenishment/bottle-change interval, backpressure and receiving geometry |
Compare product configurations by the roles on this page
| Product | Configuration option | Conditions to verify |
|---|---|---|
| HLD rotary-valve syringe pump | Aspiration, holding and dispensing with a rotary valve | Check source/outlet valve positions, capacity budget and replenishment timing |
| HMD solenoid-valve syringe pump | Metered aspiration and dispensing with solenoid valves | Check open/closed and power-off states, working volume and replenishment cycle |
Related operations and architecture comparisons
Sample preparation
Assign container identities and destinations to each step, and establish programs for the actual lysis and binding formulations. After an interruption, confirm whether the product has already been transferred before deciding whether to complete the step or stop.
Confirm product roles in actual operation
| Operation | Measurements and observations | Priority checks |
|---|---|---|
| Sample preparation | Use actual media to check first, middle and last doses and delivery after bottle changes; confirm received volume and timing rather than substituting commanded steps for volume | Empty sources, bubbles, nozzle droplets and backpressure changes affect first or last doses |
Record product volume, recovery and downstream results, and check sample-to-sample contamination.
For repeated dispensing from one chamber, is checking total volume enough?
Measure each received dose, especially the first and last. Even when total volume is near target, individual errors may differ. Relate replenishment, valve switching and receiving actions to each result.
Extraction and library-preparation liquid handling differ in contamination and outcome requirements. Confirm research workflows and clinical diagnostic methods separately.