Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Immunoassay analyzer
< Back

Piston pumps in immunoassay analyzers

Application guide

Focusing on fixed-probe sample and reagent metering in immunoassay analyzers, this page explains piston-pump location, coordination and configuration conditions, linking to related operating and validation tasks.

Applicable architecture

This page focuses on immunoassays with solid-phase separation washing, using magnetic particles as the bead-washing example. Confirm whether the specific assay requires separation washing and what solid phase it uses. Other architectures require branches defined by the actual instrument workflow. Candidate pumps and valves do not represent complete-instrument analytical performance.[4]

Application guide

Immunoassay analyzer

Immunoassay analyzer
Diaphragm pumps in immunoassay analyzers
Fluid supply and waste removal for magnetic solid-phase washing
Bubble and pressure sensing in immunoassay analyzers
Tubing and fittings in immunoassay analyzers
Fluidic valves in immunoassay analyzers
Isolation of reagent and wash-fluid paths
Piston pumps in immunoassay analyzers
Sample addition with a fixed probe
Sampling and wash probes in immunoassay analyzers
Syringe pumps in immunoassay analyzers
Multiple-reagent aspiration and dispensing

On this page

What do these products do in this instrument?Distinguish piston displacement from actual received volumeCombine metering, valve positions and container actions into a complete cycleTranslate actual branch conditions into configuration requirementsCompare product configurations by the roles on this pageExplore specific operating and validation tasksConfirm product roles in actual operationAre piston pumps of equal capacity directly interchangeable?ReferencesRelated guides

01What do these products do in this instrument?

Piston pumps: fixed-probe sample and reagent metering

Drive samples or reagents into reaction vessels in fixed-probe metering architectures. Coordinate probe motion and cleaning, distinguishing actual added volume, tip droplets and between-sample carryover.

EA, SM and TM provide options for comparing capacity, wetted configurations and installation space. Verify actual doses from working stroke and the complete wetted path; check system-liquid drive and direct sample contact with the pump head separately.

02Distinguish piston displacement from actual received volume

Piston pumps use controlled displacement for discrete aspiration and dispensing. Sample may enter the head directly or remain at the probe tip while system liquid transmits displacement; wetted coverage, priming and cleaning differ. Map aspiration, holding, dispensing and replenishment paths first, then relate stroke to probe-tip delivery.

Further technical reading

  • What Is a Precision Piston Pump? Working Principle, Benefits and Applications
  • Piston Pump Accuracy, Repeatability and Resolution Explained

03Combine metering, valve positions and container actions into a complete cycle

Delivery requires coordination of metering drive, source switching and receiving-end actions. Usable capacity must cover working volume and other cycle displacements, with replenishment in permitted windows. Check tip droplets, bubbles and shared residual liquid through the complete path; remeasure received volume after changing pump heads, tubing lengths or interfaces.

04Translate actual branch conditions into configuration requirements

Piston-pump configuration conditions in immunoassay analyzers
ConditionTo confirm
Working strokeSet working sample or reagent doses for the fixed-probe architecture, including priming and replenishment displacement within the reaction schedule.
Liquid contact and pressureMap the sample-contact or system-liquid path, probe interface and pressure states; identify surfaces included in the probe-wash sequence.
Delivery and recoveryCompare received volume and probe droplets through aspiration, delivery and washing, including the next container after a concentrated sample.
Operations supported in this instrument
OperationSpecific operating conditions
Sample addition with a fixed probeMethod-permitted sample, working dose, source level, fixed-probe geometry, liquid-contact architecture and the complete sampling/washing sequence

05Compare product configurations by the roles on this page

Configuration options for the product roles on this page
ProductConfiguration optionConditions to verify
EA standard piston pumpFinite-stroke meteringCompare configurations by routine dose, total cycle displacement and wetted materials
SM miniature piston pumpDiscrete metering in compact installationsCheck installation space first, then validate working doses and actual accuracy
TM ultra-miniature piston pumpDiscrete metering in limited spaceCheck installation space and configuration before measuring actual delivery; an ultra-miniature design does not imply a smaller reliable dose
  • EA standard piston pump

    EA standard piston pump

    View Details
  • SM miniature piston pump

    SM miniature piston pump

    View Details
  • TM ultra-miniature piston pump

    TM ultra-miniature piston pump

    View Details

06Explore specific operating and validation tasks

Sample addition with a fixed probe

Sample-volume error affects reaction ratios, while external probe droplets may contaminate the next container. Combine aspiration, delivery and probe washing into a complete sequence and compare recovery after concentrated samples.

  • Sample addition with a fixed probe

07Confirm product roles in actual operation

Operating-point validation and fault localization
OperationMeasurements and observationsPriority checks
Sample addition with a fixed probeMeasure sample received in the reaction vessel, then assess carryover through the fixed-probe wash sequence. Check first delivery after idle and recovery after concentrated samples.Air aspiration, leakage, probe-wall contact or retained droplets can affect received volume and contaminate the next container; investigate dose error and carryover separately.

Verify actual sample or reagent addition through the fixed probe and assess carryover after washing with the assay. Magnetic capture and bead-wash performance belong to the separate washing operation.

08Are piston pumps of equal capacity directly interchangeable?

Also check wetted materials, interfaces, mounting, drive and pressure conditions. After replacement, recheck delivery, cleaning and first-dose recovery with the original sequence and actual liquids. Capacity is only one comparison criterion.

Pump flow alone does not prove adequate washing; validate magnetic capture, mixing and aspiration position across the whole station.

References

View references (4)
  1. [1]FOREACH EA standard piston pump: configurations and specifications
  2. [2]FOREACH SM miniature piston pump: configurations and specifications
  3. [3]FOREACH TM ultra-miniature piston pump: configurations and specifications
  4. [4]Beckman Coulter Access: an assay example of magnetic solid-phase separation, washing and substrate detection

Related guides

  • Back to the immunoassay analyzer fluidics overview
  • Sample addition with a fixed probe

Related Products

EA Standard piston pumps

EA Standard piston pumps

SM Miniature piston pumps

SM Miniature piston pumps

TM Ultra-compact piston pumps

TM Ultra-compact piston pumps

Related Technical Articles

What Is a Precision Piston Pump? Working Principle, Benefits and Applications

Learn how precision piston pumps work, where they are used and how to select capacity, performance, materials, valves and controls for an OEM fluidic system.

2026-06-19View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

Discuss Your IVD Fluidic Requirements

Share the instrument type, sample and reagent properties, working volumes, cycle timing, fluid path and acceptance criteria so we can discuss suitable components and prototype evaluation.