Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Immunoassay analyzer
< Back

Immunoassay analyzers: fixed-probe sample addition

Application guide

Fixed-probe sample addition is a specific liquid-handling operation in immunoassay analyzers. This page explains product responsibilities, sequence, actual conditions and recovery requirements. Establish the architecture before comparing configurations.

Applicable architecture

This page focuses on immunoassays with solid-phase separation washing, using magnetic particles as the bead-washing example. Confirm whether the specific assay requires separation washing and what solid phase it uses. Other architectures require branches defined by the actual instrument workflow. Candidate pumps and valves do not represent complete-instrument analytical performance.[4]

Application guide

Immunoassay analyzer

Immunoassay analyzer
Diaphragm pumps in immunoassay analyzers
Fluid supply and waste removal for magnetic solid-phase washing
Bubble and pressure sensing in immunoassay analyzers
Tubing and fittings in immunoassay analyzers
Fluidic valves in immunoassay analyzers
Isolation of reagent and wash-fluid paths
Piston pumps in immunoassay analyzers
Sample addition with a fixed probe
Sampling and wash probes in immunoassay analyzers
Syringe pumps in immunoassay analyzers
Multiple-reagent aspiration and dispensing

On this page

Where fixed-probe sample addition occurs in the instrumentProduct responsibilities in this operationOrganize the sequence by actual architectureDefine operating conditions for this taskCompare product configurations by the roles on this pageValidate operation results at the receiverLocate abnormalities and restore the next operationReferencesRelated guides

01Where fixed-probe sample addition occurs in the instrument

This task covers sample addition with fixed sampling probes. Piston metering coordinates probe position, valves and receiving cuvettes for aspiration and delivery. Distinguish direct sample contact from system-liquid isolation before checking the sample's actual wetted path.

Set separate sequences for sample aspiration, delivery and internal/external probe washing, covering low remaining volume, first doses and high/low concentration changes. Instruments using disposable tips should refer to the air-displacement topic for tip sealing, changes and well rules.

02Product responsibilities in this operation

Piston drive creates controlled displacement, working with valve positions and the receiver to deliver this volume. Check working stroke and actual received volume separately.

For this operation, confirm: Sample identity and assay-permitted condition; direct contact or system-liquid drive; remaining source volume; probe immersion; working dose; receiving cuvette and probe-wash coverage.

03Organize the sequence by actual architecture

  1. For the fixed-probe architecture discussed here, confirm sample identity, specified dose, wetted coverage at the tip and receiving reaction vessel.
  2. Coordinate probe position and metering drive for aspiration and addition, recording tip droplets, actual delivery and affected reaction cuvettes.
  3. Clean fixed probes and shared segments to meet between-sample contamination requirements, validating received volume and contamination in the next sample separately.

04Define operating conditions for this task

Conditions, results and abnormality records for this operation
To defineDetails
Actual operating conditionsSample identity and assay-permitted condition; direct contact or system-liquid drive; remaining source volume; probe immersion; working dose; receiving cuvette and probe-wash coverage
Acceptance targetMeasure the sample received by the identified reaction cuvette, covering routine and limiting working volumes, low source volume and the first aspiration after idle. Keep actual reception separate from piston displacement.
Fault localizationAir aspiration, probe-wall contact, obstruction or leakage can leave an incomplete sample dose. Droplets and shared-path residue can contaminate the next sample even when the current cuvette receives the intended volume.

Further technical reading

  • What Is a Precision Piston Pump? Working Principle, Benefits and Applications
  • Piston Pump Accuracy, Repeatability and Resolution Explained

05Compare product configurations by the roles on this page

Configuration options for the product roles on this page
ProductConfiguration optionConditions to verify
EA standard piston pumpFinite-stroke meteringCompare configurations by routine dose, total cycle displacement and wetted materials
SM miniature piston pumpDiscrete metering in compact installationsCheck installation space first, then validate working doses and actual accuracy
TM ultra-miniature piston pumpDiscrete metering in limited spaceCheck installation space and configuration before measuring actual delivery; an ultra-miniature design does not imply a smaller reliable dose
  • EA standard piston pump

    EA standard piston pump

    View Details
  • SM miniature piston pump

    SM miniature piston pump

    View Details
  • TM ultra-miniature piston pump

    TM ultra-miniature piston pump

    View Details

06Validate operation results at the receiver

Measure the sample received by the identified reaction cuvette, covering routine and limiting working volumes, low source volume and the first aspiration after idle. Keep actual reception separate from piston displacement.

After internal and external probe washing, assess sample-to-sample residue with the method-relevant high/low sequence or blanks. Confirm sample identity, received volume and contamination results separately.

07Locate abnormalities and restore the next operation

Air aspiration, probe-wall contact, obstruction or leakage can leave an incomplete sample dose. Droplets and shared-path residue can contaminate the next sample even when the current cuvette receives the intended volume.

Hold the affected sample and cuvette, establish what was aspirated and delivered, and check probe position, the liquid column and wash state. Resample or flag the affected reaction according to the method, then verify the next sample after recovery.

This page covers fixed-probe sample addition. Disposable-tip transfer uses a different consumable sequence, and sample-volume checks do not establish solid-phase washing or complete-assay performance.

References

View references (4)
  1. [1]FOREACH EA standard piston pump: configurations and specifications
  2. [2]FOREACH SM miniature piston pump: configurations and specifications
  3. [3]FOREACH TM ultra-miniature piston pump: configurations and specifications
  4. [4]Beckman Coulter Access: an assay example of magnetic solid-phase separation, washing and substrate detection

Related guides

  • Back to the immunoassay analyzer fluidics overview
  • Piston pumps in immunoassay analyzers

Related Products

EA Standard piston pumps

EA Standard piston pumps

SM Miniature piston pumps

SM Miniature piston pumps

TM Ultra-compact piston pumps

TM Ultra-compact piston pumps

Related Technical Articles

What Is a Precision Piston Pump? Working Principle, Benefits and Applications

Learn how precision piston pumps work, where they are used and how to select capacity, performance, materials, valves and controls for an OEM fluidic system.

2026-06-19View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details