Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Immunoassay analyzer
< Back

Chemiluminescence immunoassay fluidics: dispensing, magnetic separation washing and background control

Application guide

Magnetic-particle chemiluminescence immunoassays must dispense the required sample and reagent volumes, retain the effective solid phase during separation, remove unbound components, then trigger and read the signal according to the assay. Consistent dispensing volume addresses liquid delivery; bead distribution, residual liquid after washing and reaction conditions also affect final performance.

This article focuses on magnetic separation and washing inside reaction cuvettes. Other immunoassay architectures require washing and reading locations to be defined separately. For example, some electrochemiluminescence methods aspirate the reaction mixture into a measuring cell, capture particles on an electrode and remove unbound components; their flow paths cannot simply use the same arrangement as a cuvette washing station.[1][3]

Scope

Magnetic-particle immunoassays primarily using in-cuvette magnetic separation washing; substrate, trigger reagent, magnetic capture and reading operations depend on the specific assay.

Application guide

Immunoassay analyzer

Immunoassay analyzer
Diaphragm pumps in immunoassay analyzers
Fluid supply and waste removal for magnetic solid-phase washing
Bubble and pressure sensing in immunoassay analyzers
Tubing and fittings in immunoassay analyzers
Fluidic valves in immunoassay analyzers
Isolation of reagent and wash-fluid paths
Piston pumps in immunoassay analyzers
Sample addition with a fixed probe
Sampling and wash probes in immunoassay analyzers
Syringe pumps in immunoassay analyzers
Multiple-reagent aspiration and dispensing

On this page

Which branches make up an immunoassay analyzer fluidic system?Common product roles and application guidesOperating conditions to confirm at each stageSample addition: check both volume and between-sample contaminationMultiple-reagent dispensing: bead resuspension and reagent differencesSeparation washing: retain beads and remove unbound componentsFlow-path selection: isolate substrate and cleaning-fluid sourcesFluidic monitoring: associate signals with dispensing and washing phasesExplore configurations through product application guidesValidate the complete cycle through application resultsCommon selection and recovery questionsReferences

01Which branches make up an immunoassay analyzer fluidic system?

Sample and reagent addition→Reaction and incubation→Magnetic capture and washing→Assay-specific signal triggering→Reading and waste removal
Liquid-handling task sequence

Arrows indicate task relationships, not that every component is connected in series in one tube. Check valve positions, pump wetted boundaries and operating states for each branch.

Sample, reaction reagent, wash-fluid dispensing and aspiration, substrate or trigger reagent, and waste removal each serve different tasks. Substrates and trigger reagents are options for different luminescence methods; do not assume every instrument uses both.

In-cuvette magnetic washing requires magnetic capture, dispensing, any necessary resuspension and subsequent aspiration to work together at the same station. First define the assay steps and bead-retention requirements, then arrange aspiration positions and operating windows. Extending or deepening aspiration to reduce residual liquid may increase bead loss, so both must be evaluated together.

02Common product roles and application guides

Metering, wash-fluid supply and waste removal are three different loads. Defining them separately helps avoid using a supply pump's rated flow rate as a substitute for wash volume per cuvette.

Common product roles and application guides
ComponentRole in this architectureConfiguration priorities
Piston pump / syringe pumpMetered aspiration and dispensing of sample, reagent or system liquidActual dose, liquid-contact arrangement, aspiration/dispensing speed and auxiliary strokes
Liquid diaphragm pumpWash-fluid supply and replenishment of washing stations or reservoirsConcurrent supply and installed flow resistance; the dose per cuvette is determined separately by a metering unit or validated valve control
Gas–liquid diaphragm pumpAspiration of aerated waste liquid from washing stations and reaction cuvettesPeak incoming liquid, aspiration window and liquid accumulation in tubing
Rotary valve / solenoid valveLiquid-source selection, branch switching and isolationShared volume, cross-contamination, closed state and backflow paths
Sampling probe / wash probeSample and reagent delivery; magnetic-wash dispensing and aspirationProbe-tip position, bead location, adhering droplets and residual liquid
Bubble and pressure detection, tubing and fittingsStatus signals and wetted-path connectionsPhase-specific baselines, materials, hold-up and light protection
  • View the IVD product–task matrix
  • Piston pumps in immunoassay analyzers
  • Syringe pumps in immunoassay analyzers
  • Diaphragm pumps in immunoassay analyzers
  • Fluidic valves in immunoassay analyzers
  • Sampling and wash probes in immunoassay analyzers
  • Bubble and pressure sensing in immunoassay analyzers
  • Tubing and fittings in immunoassay analyzers

03Operating conditions to confirm at each stage

When providing the following inputs for a configuration discussion, list each reagent separately so that bead suspensions and clear buffers are not treated as the same fluid.

Operating conditions to confirm at each stage
StageSpecific inputsRelated decision
Sample additionPermitted sample types, dose, low remaining volume, fixed probe or tipLiquid-contact arrangement, capacity, cleaning or consumables approach
Multiple-reagent dispensingDose of each reagent, bead resuspension, standby time and liquid pickup methodBottle-by-bottle aspiration or dedicated branches; mixing and first-dose sequence
Magnetic washing stationMagnetic capture position, dispensing volume and timing for each wash, residual-liquid and bead-retention targetsDispensing metering, aspiration positioning, waste removal and concurrent capacity
Luminescence pathwayAssay substrate or trigger reagent, temperature and light protection, interval from dispensing to readingDedicated or shared channels, materials and timing
Status and recoveryMeasurable signals, abnormal steps and handling of affected reactionsSensor locations, event association and recovery criteria

04Sample addition: check both volume and between-sample contamination

A reusable-probe design must combine sample aspiration, delivery into the cuvette and internal/external probe washing into a complete cycle, observing volume errors and low or negative results following a high sample separately. System-liquid drive also requires evaluation of dilution at the sample interface; with direct liquid contact, include the pump chamber and seals in the cleaning path.

If isolation between samples is the priority, a disposable-tip approach can be evaluated. Siemens Centaur CP uses sample tips while reagents are still aspirated with a separate probe. This choice reduces cleaning of reused sample-contact surfaces but adds consumables, loading/unloading and inventory management. Tip sealing, assembly and small-volume dispensing still require validation in the actual sequence.[1]

  • Sample addition with a fixed probe

Further technical reading

  • Piston Pump Accuracy, Repeatability and Resolution Explained

05Multiple-reagent dispensing: bead resuspension and reagent differences

Consistent aspiration volume of a bead suspension does not mean that each cuvette receives the same number of particles. After standby, first restore dispersion as specified for the reagent, then aspirate within the defined time window. For example, Centaur XPT uses an onboard reagent-pack rocking arrangement. Mixing must balance dispersion against foam formation; longer mixing and waiting consume cycle time.[2]

When a reusable probe aspirates from successive bottles, focus on pickup timing, aspiration height and cleaning between reagents. Dedicated tubing for suspensions also requires evaluation of hold-up, sedimentation and restart. For reagents that settle readily, prioritize a shorter interval from resuspension to delivery and fewer unnecessary hold-up sections; increasing pump stroke alone cannot compensate for concentration changes.

Confirm wetted materials and storage and operating conditions separately for each reagent formulation. Capacity selection must accommodate routine doses and auxiliary strokes; for batch dispensing, compare liquid and bead delivery in the first, middle and last doses and after standby recovery.

  • Multiple-reagent aspiration and dispensing

Further technical reading

  • What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

06Separation washing: retain beads and remove unbound components

During in-cuvette magnetic washing, the magnetic field can collect particles in a specific area before the liquid phase is aspirated, wash fluid added and the process repeated as required by the assay. The official Centaur CP hardware description illustrates this station arrangement. Define the magnetic-field position, cuvette geometry and aspiration-probe position together.[1]

The wash-fluid supply pump provides liquid to the station; the volume per cuvette can be metered by a metering pump or a validated valve-control unit operating under controlled pressure. When several stations dispense simultaneously, supply-pressure fluctuations may alter delivery, so measure each cuvette's volume under concurrent operation. Select waste removal for peak incoming liquid and the aspiration window.

If too much liquid remains after washing, first check aspiration height, gas–liquid conditions and accumulated waste liquid. If bead loss is excessive, first check capture time, aspiration position and the aspiration process. Additional wash cycles may improve removal of unbound components but increase liquid consumption, time and cumulative bead loss. Abnormal background also calls for checking reagent condition, nonspecific binding and the reading system; washing is not always the cause.

  • Fluid supply and waste removal for magnetic solid-phase washing

Further technical reading

  • What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?
  • IVD Waste Aspiration: Liquid Pump or Vacuum Pump? Direct vs Indirect Aspiration

07Flow-path selection: isolate substrate and cleaning-fluid sources

Select wetted-path materials, light protection and temperature conditions for the specific substrate or trigger chemistry. If valves or tubing are shared with other liquids, residue may raise background or suppress signal. Evaluate blanks and useful signal using the worst-case switching sequence rather than relying only on water flushing.

For liquids sensitive to residue or costly to clean, evaluate dedicated metering and tubing first, accepting additional components, retained liquid and maintenance points. Shared paths require defined valve positions, flushing destinations and first-dose recovery. Some electrochemiluminescence workflows capture particles and trigger readings in a measuring cell; in that case, include measuring-cell fluid renewal in the fluidic design.[3]

  • Isolation of reagent and wash-fluid paths

Further technical reading

  • Rotary Valve Selection Guide: From Fluid Paths to System Validation

08Fluidic monitoring: associate signals with dispensing and washing phases

Insufficient liquid during reagent addition affects reaction ratios; during washing it affects the wash process; abnormal waste aspiration may leave excessive residual liquid. Bubble and pressure sensors provide local signals whose meaning depends on the current station, valve position, wash cycle and normal-operation baseline.

The instrument control system records event times and associates them with samples, reaction cuvettes and completed steps. Label commanded volume, program-estimated volume and measured delivered volume separately. Without measurement evidence, these values cannot establish how much liquid is already in a cuvette. After recovery, use assay-specific criteria to decide whether to continue; if partial dispensing or washing status is uncertain, retain the exception flag and arrange a repeat test.

  • Bubble and pressure sensing in immunoassay analyzers

09Explore configurations through product application guides

Before discussing a configuration with FOREACH, clarify the following trade-offs, then compare them with the formal specifications and operating-point data for specific models.

When several metering units must handle liquids separately, evaluate HMD multichannel configurations combining glass syringes and solenoid valves. For sequential dispensing from multiple liquid sources, compare HLD configurations combining a glass syringe and multiport rotary valve. HLD ports select the flow path and do not constitute independent metering channels. Whether bead and luminescence reagents can share a path still depends on residue and assay results.[4][5]

  • Sample and reagent metering: compare piston and syringe pumps by actual dose, wetted path, auxiliary strokes and speed sequence; pump type alone does not establish small-dose accuracy.
  • Wash-fluid supply and station metering: separate supply, per-cuvette metering, valve control and waste removal, then check volume and pressure changes during simultaneous operation of multiple stations.
  • Bead branch: coordinate probes, valves and tubing to reduce hold-up, using aspiration and recovery sequences suited to the actual suspension.
  • Substrate or trigger reagent: prioritize comparison of dedicated channels versus shared-path flushing in terms of background, useful signal, liquid use and maintenance cost.
  • Monitoring and connections: choose sensing locations for the abnormalities to be detected, and verify wetted materials, tubing dead volume and any required light protection.
  • Organize configurations using the IVD product–task matrix
  • Piston pumps in immunoassay analyzers
  • Syringe pumps in immunoassay analyzers
  • Diaphragm pumps in immunoassay analyzers
  • Fluidic valves in immunoassay analyzers
  • Sampling and wash probes in immunoassay analyzers
  • Bubble and pressure sensing in immunoassay analyzers
  • Tubing and fittings in immunoassay analyzers

10Validate the complete cycle through application results

Evaluate delivery, bead retention, residual liquid and signal together for the specific assay. If lower background is accompanied by lower useful signal, check bead loss or signal suppression; a single blank value is insufficient to evaluate the configuration.

Validate the complete cycle through application results
StageWhat to measurePriority checks
Sample additionVolume received in the cuvette; low or negative results following a high sampleAir aspiration, liquid on outer surfaces, tip sealing and cleaning
Multiple reagentsActual liquid volume, first/middle/last doses, and consistency of particle delivery after resuspensionSedimentation, foam, low remaining volume and residence time
Magnetic washingDispensed volume per wash, residual liquid, bead loss and assay resultsInsufficient capture, misplaced aspiration, concurrent supply and accumulated waste liquid
Luminescence pathwayBlank and useful signal after switching; interval from dispensing to readingShared-path residue, reagent condition, timing and detection system
Recovery from abnormalitiesDetection and false alarms, station traceability, first cycle after recoveryBaselines, wash-cycle association and uncertain partial delivery
  • View common operating-point and recovery validation methods

11Common selection and recovery questions

Why can background remain high even when every dispensing volume is consistent?

Consistent volume verifies only part of delivery. Compare bead dispersion, capture and loss, residual liquid after washing and waste-removal conditions in sequence, then check reagent condition, nonspecific binding and the reading system. Locate the issue through staged experiments before adjusting the process, rather than addressing every problem by increasing wash volume.

Can bead reagent, substrate and wash fluid share one path?

This depends on the actual formulations, path and residue permitted by the assay. Dedicated channels are easier to manage if deposition or residue makes flushing costly. Shared channels require validation of residue, blanks, useful signal and the first dose after standby under the worst-case sequence. A multiport valve alone does not prove that these liquids can share a path.

References

View references (5)
  1. [1]Siemens Healthineers Academy — ADVIA Centaur CP Immunoassay System Hardware Overview
  2. [2]Siemens Healthineers — ADVIA Centaur XPT,Reagent Handling
  3. [3]Roche — Elecsys proBNP II Method Sheet,07027664500 V4.0,Test principle
  4. [4]FOREACH | HMD solenoid-valve syringe pump series
  5. [5]FOREACH | HLD rotary-valve syringe pump series

Related Products

HLD Rotary-valve syringe pumps

HLD Rotary-valve syringe pumps

EA Standard piston pumps

EA Standard piston pumps

DPL30 Liquid diaphragm pumps

DPL30 Liquid diaphragm pumps

DPGL800 Gas–liquid diaphragm pumps

DPGL800 Gas–liquid diaphragm pumps

Related Technical Articles

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

A programmable syringe pump uses a motor to control syringe plunger displacement and speed, executing aspiration, dispensing, pauses and repeated actions through commands. With a routing valve, it can connect reagent reservoirs, reaction vessels and wash paths into an automated dispensing process. Instrument development therefore needs to define volume, fluid connections and timing together.

2026-09-28View Details

What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?

Cleaning pump, wash pump and rinse pump usually name different duties within an IVD cleaning sequence, not three fixed pump mechanisms. Selection starts with wash volume, cycle rate, line resistance, wetted materials, and whether the duty is liquid supply or waste aspiration.

2026-09-03View Details

IVD Waste Aspiration: Liquid Pump or Vacuum Pump? Direct vs Indirect Aspiration

Probe washing, reaction-cup cleaning, and routine waste removal in IVD instruments are all described as “waste aspiration.” Yet whether waste passes directly through a liquid pump or first enters a collection bottle while a vacuum pump creates negative pressure leads to entirely different selection logic.

2026-08-09View Details

Discuss Your IVD Fluidic Requirements

Share the instrument type, sample and reagent properties, working volumes, cycle timing, fluid path and acceptance criteria so we can discuss suitable components and prototype evaluation.

Discuss your application