Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Coagulation analyzer
< Back

Coagulation analyzer fluidics: organize liquid delivery around reaction initiation

Application guide

Coagulation fluidics must deliver metered plasma and reagent volumes to the reaction site while coordinating liquid arrival, mixing, temperature and detection. In clotting-time tests, the dispensing action that starts the reaction is especially important. The controller command, liquid leaving the probe and liquid contacting the reaction mixture are distinct events whose relationship must be established.

One instrument may run clotting, chromogenic and immunoturbidimetric assays, with different addition sequences, incubation periods and acquisition windows. Define reaction timing for the method before discussing pump capacity, probe-tip structure and path length, so component response becomes repeatable liquid delivery.[1]

Scope

For development of coagulation analyzers using plasma samples; detection start, temperature and mixing requirements are jointly determined by the reagent method and instrument sequence.

Application guide

Coagulation analyzer

Coagulation analyzer
Diaphragm pumps in coagulation analyzers
Cleaning reused paths and recovering the first dose
Bubble and pressure sensing in coagulation analyzers
Tubing and fittings in coagulation analyzers
Multiport rotary valves in coagulation analyzers
Switching reagent sources and reaction paths
Piston pumps in coagulation analyzers
Assay-specified sample delivery
Sampling probes in coagulation analyzers
Syringe pumps in coagulation analyzers
Reagent arrival and reaction timing

On this page

Which branches make up a coagulation analyzer fluidic system?Common product roles and application guidesOperating conditions to confirm at each stageSample delivery: control receiving position and residueReagent addition: use actual arrival as the timing referencePath switching: align valve positions with reaction timingCleaning: protect the next reaction's starting conditionsState monitoring: identify affected reaction cuvettesExplore configurations through product application guidesValidate the complete cycle through application resultsCommon selection and recovery questionsReferences

01Which branches make up a coagulation analyzer fluidic system?

Metered plasma delivery→Reagent preparation and prewarming→Sequential addition→Mixing and detection timing→Probe washing and first-dose recovery
Liquid-handling task sequence

Arrows indicate task relationships, not that every component is connected in series in one tube. Check valve positions, pump wetted boundaries and operating states for each branch.

The main branches handle plasma dispensing, reagent addition, probe washing and waste removal. Separate reagent probes, heated sections or dedicated channels depend on the assay mix, storage conditions and test cycle time.

Clot detection may identify clot formation optically or mechanically; chromogenic methods read color development and immunoturbidimetric methods read the turbidity change of the corresponding reaction. Optical detection requires attention to bubbles and liquid-surface effects on the optical path; mechanical clot detection still requires consistent dispensing, mixing and reaction conditions. Multiple detection modules in one instrument do not justify using the same trigger time for all assays.[1][2]

02Common product roles and application guides

Configure components backward from the actual receiving location: establish probe-tip delivery first, then metering, routing, prewarming and cleaning. Heating and result interpretation are instrument functions that must be coordinated with pump and valve actions.

Common product roles and application guides
ProductTask in a coagulation analyzer
Piston pump / syringe pumpDrive metered aspiration and dispensing of plasma and reagents, matching actual doses and auxiliary strokes
Liquid diaphragm pumpSupply liquid to clean inner and outer probe surfaces, accounting for wash-station resistance and simultaneous demand
Multiport rotary valves / solenoid valvesSelect liquid sources and cleaning paths, control on/off states and coordinate metering and prewarming timing
Sampling and wash probesPerform sampling and in-cuvette delivery, coordinated with positioning, cleaning and instrument thermal design
Bubble and pressure detectionOutput abnormality signals that the instrument associates with dispensing steps and reaction cuvettes
Tubing and fittingsConnect sources to probe tips, controlling resistance, retained volume and assembly consistency
  • Piston pumps in coagulation analyzers
  • Syringe pumps in coagulation analyzers
  • Diaphragm pumps in coagulation analyzers
  • Multiport rotary valves in coagulation analyzers
  • Sampling probes in coagulation analyzers
  • Bubble and pressure sensing in coagulation analyzers
  • Tubing and fittings in coagulation analyzers

03Operating conditions to confirm at each stage

Draw one test as a timeline, marking aspiration, waiting, prewarming, dispensing, mixing and acquisition separately. Storage temperature, heater setpoint and liquid arrival temperature are different conditions.

Operating conditions to confirm at each stage
StageConditions determining configuration
Plasma dispensingDose range, receiving position at the probe tip, whether sample enters the pump chamber, isolation interface
Reagent additionEach reagent's dose and order, assay-specified arrival temperature and permissible addition window
Detection timingTrigger event, mixing, preincubation and acquisition window for clotting, chromogenic or immunoturbidimetric assays
Path switchingValve position, shared-segment volume, flushing consumption and first-dose recovery time after switching
Cleaning and monitoringInner/outer surface coverage, residual wash fluid and confirmable action status at alarm

04Sample delivery: control receiving position and residue

Small plasma doses require attention to droplets at the probe tip, liquid adhering to the cuvette wall and the system-liquid interface. Measure received volume using the actual cuvette and probe position, then assess whether the liquid reaches the reaction zone promptly. Faster dispensing may reduce hanging droplets but increase splashing and foam; choose the speed profile by end-point behavior.

With system-liquid drive, examine the isolation interface and sample dilution. With direct contact, include the pump chamber, valves and connectors in the cleaning path. Capacity must cover sample, isolation segments and necessary auxiliary actions while retaining useful stroke for routine small doses; resolve the large-capacity versus small-dose trade-off using probe-tip delivery data.

  • Assay-specified sample delivery

05Reagent addition: use actual arrival as the timing reference

Use suitable development observations to establish the timing between the control signal, liquid leaving the probe, contact with liquid in the cuvette and completion of dispensing. For clotting assays, associate the initiation event with timing as defined by the method; for chromogenic and immunoturbidimetric assays, schedule acquisition by their incubation and reading rules. Completion of pump-motor movement is only one control event.

The method determines whether and where reagents are prewarmed and the permitted temperature at arrival. Longer heating residence may bring temperature closer to target but changes cycle time and reagent exposure. Short tubing reduces retained liquid and transfer distance, yet heat loss at the probe tip still needs checking. After changing probes, tubing length or aspiration/dispensing speed, review arrival temperature, delay and first-dose behavior together.

  • Reagent arrival and reaction timing

Further technical reading

  • Piston Pump Accuracy, Repeatability and Resolution Explained
  • What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

06Path switching: align valve positions with reaction timing

One metering unit with multiport routing can aspirate different liquids sequentially when flushing waits are acceptable. Dedicated paths reduce shared-segment flushing for residue-sensitive, frequently used or differently temperature-controlled reagents, at the cost of more metering and cleaning units.

Each valve position should have a defined source, destination and permitted action. Even after the valve reaches position, the shared segment may retain the previous liquid. Complete assay-required flushing and preparation before the dispensing that starts the reaction. Place valve switching, priming and the cuvette reaction on one timeline to identify the true throughput-limiting step.

  • Switching reagent sources and reaction paths

Further technical reading

  • Rotary Valve Selection Guide: From Fluid Paths to System Validation

07Cleaning: protect the next reaction's starting conditions

Treat internal residue, liquid on the probe exterior and residual wash fluid separately. Internal flushing addresses the wetted path; external washing covers immersed and splashed surfaces; post-cleaning drainage or recovery affects the next dose's concentration and volume. If extra wash fluid solves a local issue, also evaluate waste load and water remaining for the next dose.

Check residue effects in the assay-switching sequence, especially the first dose after cleaning. If only the first dose is abnormal, first examine flushing, tip droplets and thermal state before changing recovery. Each extra pre-dispense consumes reagent and cycle time.

  • Cleaning reused paths and recovering the first dose

Further technical reading

  • What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?

08State monitoring: identify affected reaction cuvettes

Bubble and pressure sensors provide fluidic-state signals; the instrument control system associates cuvette number, assay and dispensing step. Alarm records should retain event time, valve position, commanded volume, action completion and actual volume only when independently measured. Sending a dispensing command alone cannot confirm the volume received in the cuvette.

Use different handling rules for abnormalities before and after reaction initiation. During recovery, first establish the cuvette's stage, then decide under the method whether to continue, discard or repeat. After partial dispensing with unknown actual volume, do not simply resend the full dose to the same cuvette.

  • Bubble and pressure sensing in coagulation analyzers

09Explore configurations through product application guides

FOREACH HMD combines glass syringes with solenoid valves and can be compared for single or multiple metering channels. HLD combines a glass syringe with a multiport rotary valve for sequential routing with one metering unit. Focus on the number of metering units needed, replenishment timing and shared wetted segments, not port count alone.[3][4]

Compare piston and syringe pumps by dose, auxiliary strokes and aspiration/dispensing speed; configure cleaning pumps around actual supply and first-dose recovery. Coordinate probes and tubing with receiving position, retained volume and thermal structure, and place sensors according to the phase of the abnormality. Once a model is selected, check the design against its capacity, materials and operating-point data.

  • View the IVD product–task matrix
  • Piston pumps in coagulation analyzers
  • Syringe pumps in coagulation analyzers
  • Diaphragm pumps in coagulation analyzers
  • Multiport rotary valves in coagulation analyzers
  • Sampling probes in coagulation analyzers
  • Bubble and pressure sensing in coagulation analyzers
  • Tubing and fittings in coagulation analyzers

10Validate the complete cycle through application results

Check both delivery and assay results using the same actual cuvettes, probes, tubing and reagents. Cover routine continuous operation, standby recovery, liquid changes and the first dose after cleaning separately.

Validate the complete cycle through application results
StageActual measurementPriority areas to investigate
Plasma deliveryVolume received in the cuvette, residual droplets and between-sample carryoverAir aspiration, hanging tip droplets, cuvette-wall adhesion and isolation interface
Reagents and timingAdded volume, arrival time and variability, arrival temperature and complete reactionTrigger mapping, mixing, temperature control and speed sequence
Path switchingResidue after flushing, first-dose volume and preparation timeLiquid retained in shared segments, valve position and flushing sequence
Cleaning recoveryFirst-dose concentration or volume, assay results and liquid useInner/outer coverage, residual water and thermal state
Abnormality identificationPhase-specific detection, false alarms, response and cuvette-number associationSignal location, event time and control state
  • General validation methods and boundary operating conditions

11Common selection and recovery questions

Can completion of pump movement directly define the start of clotting?

First measure the relationship between the pump signal, liquid leaving the probe and contact in the cuvette, then define the start according to the method. A stable control offset can be included in instrument timing; excessive variability calls for addressing bubbles, hanging droplets, the path or drive first.

Does faster dispensing shorten the test cycle?

Only if actual dispensing, mixing, temperature and method-specific waiting requirements are all met. First locate the bottleneck in aspiration, prewarming, flushing or the detection window. If the method fixes the wait, shortening motor motion alone may not increase actual throughput.

References

View references (4)
  1. [1]Stago — STA Compact Max Technical Specifications (2019), page 2, Measurement
  2. [2]Stago — Group Brands Presentation
  3. [3]FOREACH | HMD solenoid-valve syringe pump series
  4. [4]FOREACH | HLD rotary-valve syringe pump series

Related Products

EA Standard piston pumps

EA Standard piston pumps

HLD Rotary-valve syringe pumps

HLD Rotary-valve syringe pumps

DPL30 Liquid diaphragm pumps

DPL30 Liquid diaphragm pumps

Related Technical Articles

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

A programmable syringe pump uses a motor to control syringe plunger displacement and speed, executing aspiration, dispensing, pauses and repeated actions through commands. With a routing valve, it can connect reagent reservoirs, reaction vessels and wash paths into an automated dispensing process. Instrument development therefore needs to define volume, fluid connections and timing together.

2026-09-28View Details

Rotary Valve Selection Guide: From Fluid Paths to System Validation

Map the connections required at each step before comparing specifications. Use the actual liquids, tubing and switching sequence to validate suitability for your instrument.

2026-09-30View Details

What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?

Cleaning pump, wash pump and rinse pump usually name different duties within an IVD cleaning sequence, not three fixed pump mechanisms. Selection starts with wash volume, cycle rate, line resistance, wetted materials, and whether the duty is liquid supply or waste aspiration.

2026-09-03View Details

Discuss Your IVD Fluidic Requirements

Share the instrument type, sample and reagent properties, working volumes, cycle timing, fluid path and acceptance criteria so we can discuss suitable components and prototype evaluation.