Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Coagulation analyzer
< Back

Coagulation analyzers: reagent arrival and reaction timing

Application guide

Reagent arrival and reaction timing is a specific liquid-handling operation in coagulation analyzers. This page explains product responsibilities, sequence, actual conditions and recovery requirements. Establish the architecture before comparing configurations.

Applicable architecture

Coagulation instruments may run clotting, chromogenic or immunoturbidimetric methods. This page discusses coordination of liquid delivery and reaction timing. Define initiation, preincubation and reading rules by method rather than reducing every assay to the same clotting endpoint.[3]

Application guide

Coagulation analyzer

Coagulation analyzer
Diaphragm pumps in coagulation analyzers
Cleaning reused paths and recovering the first dose
Bubble and pressure sensing in coagulation analyzers
Tubing and fittings in coagulation analyzers
Multiport rotary valves in coagulation analyzers
Switching reagent sources and reaction paths
Piston pumps in coagulation analyzers
Assay-specified sample delivery
Sampling probes in coagulation analyzers
Syringe pumps in coagulation analyzers
Reagent arrival and reaction timing

On this page

Where reagent arrival and reaction timing occur in the instrumentProduct responsibilities in this operationOrganize the sequence by actual architectureDefine operating conditions for this taskCompare product configurations by the roles on this pageValidate operation results at the receiverLocate abnormalities and restore the next operationReferencesRelated guides

01Where reagent arrival and reaction timing occur in the instrument

Use actual reagent arrival at the reaction site as the timing reference for dispensing, mixing, temperature control and measurement windows. Validate probe movement, transfer delay and replenishment recovery; completion of motor movement is not the reaction start.

Measure dispensing delay and its variability before relating control events to detection timing. Recheck the delay after changing tubing length or probes.

02Product responsibilities in this operation

Syringe drive and valve positions jointly organize aspiration, holding and dispensing. Record working volume, remaining inventory and replenishment timing according to the sequence.

For this task in coagulation analyzers, confirm working volume, addition order, liquid-arrival delay, mixing wait and detection trigger.

03Organize the sequence by actual architecture

  1. Confirm reagent order, preincubation, mixing and reading windows for the current clotting, chromogenic, immunoturbidimetric or other method.
  2. Verify reagent source and receiving cuvette, perform metered delivery and record pump action, liquid leaving the probe and actual arrival separately.
  3. Relate actual arrival and mixing to the assay-defined trigger event; pump-motion completion cannot serve as the universal reaction start for all methods.
  4. After changing tubing length, probes or speed, recheck volume, arrival delay and complete-assay results before evaluating the new cycle time.

04Define operating conditions for this task

Conditions, results and abnormality records for this operation
To defineDetails
Actual operating conditionsWorking volume, addition order, liquid-arrival delay, mixing wait and detection trigger
Acceptance targetRecord the pump command, probe-tip outflow and actual reagent arrival at the cuvette separately. Measure delivered volume and arrival-delay variation before relating them to the method-defined mixing, initiation and reading events.
Fault localizationA delayed droplet, incomplete dose or insufficient mixing can shift the reaction conditions while the control trigger remains unchanged. Command timing therefore needs to be reconciled with actual liquid contact.

Further technical reading

  • What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation
  • Piston Pump Accuracy, Repeatability and Resolution Explained

05Compare product configurations by the roles on this page

Configuration options for the product roles on this page
ProductConfiguration optionConditions to verify
HLD rotary-valve syringe pumpAspiration, holding and dispensing with a rotary valveCheck source/outlet valve positions, capacity budget and replenishment timing
HMD solenoid-valve syringe pumpMetered aspiration and dispensing with solenoid valvesCheck open/closed and power-off states, working volume and replenishment cycle
  • HLD rotary-valve syringe pump

    HLD rotary-valve syringe pump

    View Details
  • HMD solenoid-valve syringe pump

    HMD solenoid-valve syringe pump

    View Details

06Validate operation results at the receiver

Record the pump command, probe-tip outflow and actual reagent arrival at the cuvette separately. Measure delivered volume and arrival-delay variation before relating them to the method-defined mixing, initiation and reading events.

After a change to tubing, probes or speed, recheck the complete reaction with its required addition order and temperature conditions. Evaluate first-dose recovery after standby as well as repeated operation.

07Locate abnormalities and restore the next operation

A delayed droplet, incomplete dose or insufficient mixing can shift the reaction conditions while the control trigger remains unchanged. Command timing therefore needs to be reconciled with actual liquid contact.

Hold the affected cuvette and review which reagents arrived, their volumes and their timing. Apply the assay procedure to the interrupted reaction, restore the delivery path, and verify a new complete sequence before resuming measurements.

The assay defines its initiation and reading events. Pump or valve response time is not a universal reaction-start time.

References

View references (3)
  1. [1]FOREACH HLD rotary-valve syringe pump: configurations and specifications
  2. [2]FOREACH HMD solenoid-valve syringe pump: configurations and specifications
  3. [3]Sysmex CA-600: clotting, chromogenic and immunoturbidimetric configurations

Related guides

  • Back to the coagulation analyzer fluidics overview
  • Syringe pumps in coagulation analyzers

Related Products

HLD Rotary-valve syringe pumps

HLD Rotary-valve syringe pumps

HMD Solenoid-valve syringe pumps

HMD Solenoid-valve syringe pumps

Related Technical Articles

What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

A programmable syringe pump uses a motor to control syringe plunger displacement and speed, executing aspiration, dispensing, pauses and repeated actions through commands. With a routing valve, it can connect reagent reservoirs, reaction vessels and wash paths into an automated dispensing process. Instrument development therefore needs to define volume, fluid connections and timing together.

2026-09-28View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

Discuss Your IVD Fluidic Requirements

Share the instrument type, sample and reagent properties, working volumes, cycle timing, fluid path and acceptance criteria so we can discuss suitable components and prototype evaluation.