FOREACH
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control Series
Applications
IVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About FOREACHR&D and ManufacturingQuality System and QualificationsMilestonesFOREACH Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About FOREACH
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • FOREACH Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Products/Piston Pump/SM-500-PMMA
FOREACH 500 μL Miniature Precision Ceramic Piston Pump for Reagent Addition and Metered Diluent Dosing, with a PMMA Pump Head

FOREACH 500 μL Miniature Precision Ceramic Piston Pump for Reagent Addition and Metered Diluent Dosing, with a PMMA Pump Head

SM-500-PMMA is a 500 μL miniature precision piston pump with a PMMA head for reagent addition and metered diluent dosing. Designed for integration into small automated instruments and sampling mechanisms, it supports 1/4-28 UNF or M6 ports and 2000 full-stroke steps. Full-stroke repeatability is ≤0.5% under the specified conditions.

SM head options include PMMA, PSU, POM and PEEK; piston options include ceramic, stainless steel and PEEK. Optical sensors, valves, probes, tubing and a controller can be matched to the project. Materials and mounting are selected for the fluid and instrument.

Typical Applications:

Analyzer reagent dispensing, Automated sampling and sample dilution, Diluent and buffer dispensing, Water-analysis reagent metering

Model:This product is custom-made
Volume500 μL
Displayed pump head materialPoly(methyl methacrylate) (PMMA)
Pump head material optionsPMMA / PSU / POM / PEEK; final selection follows fluid and structural requirements.
Piston material optionsCeramic / stainless steel / PEEK; select with the seal and fluid.
Fluidic interface1/4-28 UNF / M6
Stroke12.7 mm
Recommended max speed600 RPM
Motor step angle1.8°
Lead1.27 mm
Full-stroke steps2000 Step
Accuracy at 100% strokeConfirm for the selected configuration and working volume
Repeatability at 100% stroke≤0.5% (specified full-stroke conditions)
Repeatability at 2% strokeConfirm for the selected configuration and working volume
Backlash< 1.0%
Designed life5 million cycles (pure water, room temperature, 50 kPa back pressure)
Max fluid pressure<0.30 MPa

SM-500-PMMA applications and selection requirements

SM miniature piston pumps can be configured for clinical chemistry, immunoassay, laboratory automation, sample preparation, life science instruments and water or environmental analysis. Depending on the fluidic task, selection may focus on a reagent dispensing pump, sampling pump, sample dilution pump, reaction-liquid dispensing pump, buffer dosing pump or standard-addition pump.

SM-500-PMMA focuses on reagent addition and metered diluent dosing. Select the 500 μL capacity against the working dose. If the total transfer exceeds one stroke, include refill and valve-switching time. The displayed configuration uses a PMMA head and ceramic piston. All wetted materials must be checked against the actual fluid.

Clinical chemistry and immunoassay: reagent dispensing pump

Analyzers repeatedly add programmed reagent doses to reaction units. Selection depends on consistency at the working dose and coordination between the pump, selector valve and dispensing probe.

SM-500-PMMA: Aspiration and dispense strokes can be coordinated with reagent-routing valves; wetted materials are selected against reagent composition, temperature and cleaning conditions. SM supports integration with automated sampling mechanisms; 2000 full-stroke steps provide position control for aspiration and dispensing. Full-stroke repeatability does not establish performance at smaller doses.

Laboratory automation: sampling and sample dilution pump

Automated sampling and sample preparation require separate control of sample and diluent volumes, together with management of carryover, bubbles and valve timing.

SM-500-PMMA: The pump can be configured with probes, tubing and valves for sampling and dispensing. Dilution-ratio performance also requires validation of sample volume, diluent volume and mixing. Select the 500 μL capacity against the working dose. If the total transfer exceeds one stroke, include refill and valve-switching time.

Automated liquid preparation: diluent and buffer dosing pump

Liquid-preparation and analytical modules meter diluents or buffers. Capacity selection must account for dose volume, refill cycles and instrument timing.

SM-500-PMMA: Metered additions can use a single stroke or repeated aspiration and dispense cycles. Saline buffers require shutdown flushing and residual-liquid management to address crystallization. Select the 500 μL capacity against the working dose. If the total transfer exceeds one stroke, include refill and valve-switching time.

Water and environmental analysis: reagent metering and standard-addition pump

Small water-analysis modules add defined reagent or standard volumes to measurement cells. Selection depends on the dosing range, fluid chemistry, concentration and maintenance conditions.

SM-500-PMMA: Dose capacity and fluid-path switching can be matched to the measurement sequence. The head, piston, seals and valves must be assessed together; corrosive, particulate or crystallization-prone fluids require specific validation. Select the 500 μL capacity against the working dose. If the total transfer exceeds one stroke, include refill and valve-switching time. The displayed configuration uses a PMMA head and ceramic piston. All wetted materials must be checked against the actual fluid.

Confirm the working conditions

Provide dose volume, allowed error, timing, fluid composition, temperature, back pressure and mounting space. 500 μL divided by 2000 steps gives approximately 0.25 μL per step as a nominal calculation only. Validate minimum reliable dose, carryover and dripping in the complete fluid path.

The SM specification gives an expected life of five million cycles with pure water at room temperature and 50 kPa back pressure. Saline or crystallization-prone fluids require shutdown flushing and maintenance planning.

Loading 3D model...
Preview DrawingView the technical drawing for FOREACH 500 μL Miniature Precision Ceramic Piston Pump for Reagent Addition and Metered Diluent Dosing, with a PMMA Pump Head.
Download SM PDF
Preview SM Datasheet

Frequently Asked Questions

It focuses on reagent addition and metered diluent dosing. Related selection tasks include analyzer reagent dispensing, automated sampling and sample dilution, diluent and buffer dispensing, water-analysis reagent metering. Select the 500 μL capacity against the working dose. If the total transfer exceeds one stroke, include refill and valve-switching time.

No. 500 μL is the nominal capacity. 2000 full-stroke steps give approximately 0.25 μL per step, which is not a validated minimum dose either. Fluid, backlash, bubbles, valves and tubing affect results; test at the intended working volume.

The displayed configuration uses a PMMA head and ceramic piston. All wetted materials must be checked against the actual fluid. SM head options include PMMA, PSU, POM and PEEK; piston options include ceramic, stainless steel and PEEK. Optical sensors, valves, probes, tubing and a controller can be matched to the project. Materials and mounting are selected for the fluid and instrument.

SM supports integration with automated sampling mechanisms; 2000 full-stroke steps provide position control for aspiration and dispensing. Full-stroke repeatability does not establish performance at smaller doses. See the capacity-specific technical drawing on this page for dimensions. Body dimensions exclude external fittings and tubing clearance; the selected configuration drawing governs.

The displayed configuration uses 1/4-28 UNF / M6 ports. Home-position optical sensors, valves and an ISC1000 controller can be selected; RS-232, RS-485 and CAN options depend on the chosen controller configuration. Tubing diameter, seals and mounting orientation must also be specified.

SM full-stroke repeatability is ≤0.5% under the specified conditions, not at every small dose. Provide the working volume and allowed error to confirm configuration-specific accuracy. The expected five-million-cycle life uses pure water at room temperature and 50 kPa back pressure; it is not a universal reagent-life guarantee.

Related Technical Articles

How to Select a Piston Pump Head Material

Compare PMMA, PCTG, PPS, PVDF, PP, PTFE, PEEK, POM and PSU pump-head options, and learn how to validate the complete wetted path.

2026-09-06View Details

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Precision Piston Pump? Working Principle, Benefits and Applications

Learn how precision piston pumps work, where they are used and how to select capacity, performance, materials, valves and controls for an OEM fluidic system.

2026-06-19View Details

Acceleration and Deceleration Curves for Piston-Pump Drives

Use stepper-motor torque-speed behavior to select linear, S-curve or parabolic motion profiles and prevent missed steps in piston-pump aspiration and dispensing.

2026-09-06View Details

Precision Piston Pump Backlash Compensation Guide

Understand reversal backlash, measurement methods and compensation timing, and distinguish mechanical lost motion from actual aspirated or dispensed volume.

2026-09-06View Details

Precision Piston Pump Stepper Motor Calculation and Selection

Calculate pulse frequency, microstepping, transmission ratio and load torque, then verify speed and dynamic torque with a 400 mm reciprocating-platform example.

2026-09-06View Details

Piston Pump Configured for Your Instrument

Share the target volume, fluid compatibility, port type, control method, installation space, and service-life requirements. The FOREACH engineering team can help confirm the pump configuration and wetted materials.

Submit a Custom Request