Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Clinical chemistry analyzer
< Back

Clinical chemistry analyzer fluidics: from sample metering to cuvette delivery

Application guide

Clinical chemistry analyzers add samples and method-specified reagents to reaction cuvettes, then mix, incubate and measure optically. The end of one dispensing step begins the next washing and recovery sequence: metering deviations change reaction ratios, while residue in wetted paths may affect subsequent tests.

This guide focuses on automated chemistry architectures with reusable sample probes, reagent probes and reaction cuvettes, examining sample, reagent, wash and waste branches separately. Sample types, single- or multiple-reagent workflows, addition order and reaction conditions are defined by the specific method; one R1/R2 program cannot be applied to every assay.

Scope

Automated clinical chemistry fluidics with reusable probes; disposable-tip, dry-chemistry and special reaction-cell systems require configurations appropriate to their own architectures.

Application guide

Clinical chemistry analyzer

Clinical chemistry analyzer
Diaphragm pumps in clinical chemistry analyzers
Probe/cuvette wash supply and waste removal
Bubble and pressure detection in clinical chemistry analyzers
Tubing and fittings in clinical chemistry analyzers
Multiport rotary valves in clinical chemistry analyzers
Sample, reagent and wash-path switching
Piston pumps in clinical chemistry analyzers
Metered reagent dispensingSample aspiration and transfer
Sampling and wash probes in clinical chemistry analyzers
Syringe pumps in clinical chemistry analyzers

On this page

Which branches make up clinical chemistry analyzer fluidics?Common product roles and application guidesOperating conditions to confirm at each stageSample aspiration: dose range, liquid-level detection and wall contactReagent dispensing: R1/R2 delivery and refill coordinationFluidic switching: valve positions, shared volume and flushingProbe and cuvette washing: treat inner surfaces, outer surfaces and cuvette bottoms separatelyTubing and fittings: why recheck shared volume after servicing?Level, bubble and pressure detection: linking alarms to specific testsExplore configurations through product application guidesValidate the complete cycle through application resultsCommon selection and recovery questionsReferencesRelated guides

01Which branches make up clinical chemistry analyzer fluidics?

Sample aspiration→Method-specific reagent addition→Mixing and reaction→Probe and cuvette washing→Recovery for the next test
Liquid-handling task sequence

Arrows indicate task relationships, not that every component is connected in series in one tube. Check valve positions, pump wetted boundaries and operating states for each branch.

Sample and reagent metering deliver liquid to reaction cuvettes; wash supply and waste aspiration restore wetted surfaces. Branches may share a water source or waste system, but metering, washing and waste removal require separate operating points.

Identify the liquid-contact arrangement before selecting metering hardware. With system-liquid drive, the pump mainly contacts system liquid and the probe tip contacts sample; control air gaps, interface mixing and auxiliary stroke. With direct contact, sample enters the pump chamber, so materials, cleaning and residue management must cover the chamber and seals. The two arrangements need different wetted-path assessments.

02Common product roles and application guides

Select functions by branch first, then match candidate capacity, materials and drive configurations to actual operating points. The following component roles apply to the architecture discussed here.

Common product roles and application guides
ComponentTask performedConfiguration priorities
Piston pump / syringe pumpMetered aspiration and dispensing of samples, reagents or system liquidStroke utilization at actual doses, auxiliary stroke and aspiration/dispense timing
Liquid diaphragm pumpSupply for probe washing, cuvette washing or reservoirsFlow under installed resistance and supply during concurrent demand
Gas–liquid diaphragm pumpAspiration of air-containing waste liquidAspiration height, gas–liquid mixture state and peak incoming flow
Rotary valve / solenoid valveLiquid-source selection and wash/waste on/off controlShared volume, cross-leakage and switching states
Sampling probe / wash probeSampling from containers, cuvette delivery and cleaning of inner/outer surfaces and cuvette bottomsPositioning, retained droplets, wash coverage and residual liquid
Bubble and pressure detection, tubing and fittingsProvide state signals and connect branchesInstallation position, wetted materials, dead volume and assembly consistency
  • View the IVD product–task matrix
  • Piston pumps in clinical chemistry analyzers
  • Syringe pumps in clinical chemistry analyzers
  • Diaphragm pumps in clinical chemistry analyzers
  • Multiport rotary valves in clinical chemistry analyzers
  • Sampling and wash probes in clinical chemistry analyzers
  • Bubble and pressure detection in clinical chemistry analyzers
  • Tubing and fittings in clinical chemistry analyzers

03Operating conditions to confirm at each stage

Define requirements for specific liquids, containers and actions before using capacity and flow as selection criteria.

Operating conditions to confirm at each stage
StageProject inputsSelection directly affected
Sample aspirationMethod-permitted serum, plasma or urine; dose, tube type and low remaining volumeLiquid-contact arrangement, pump capacity, probe positioning and aspiration program
Reagent dispensingActual dose of each reagent, addition order, aspiration architecture and timing windowBottle-by-bottle aspiration or dedicated branch; refill and dispensing timing
Path switchingLiquid combinations, valve positions, shared segments and allowable residueShared or independent channels, flushing and cleaning programs
Probe and cuvette washingWash-liquid composition, surfaces to clean and peak simultaneous inflowSupply, station distribution and waste-removal capacity
Monitoring and maintenanceFault-handling rules, measurable signals and allowable recovery timeDetection locations, record fields and post-maintenance checks

04Sample aspiration: dose range, liquid-level detection and wall contact

Screen capacities using minimum, routine and maximum delivered volumes, leaving stroke for required air gaps, suck-back and other actions. Larger capacity reduces refills but lowers the stroke fraction used by the same small dose. Setting resolution describes command granularity only; small-dose performance depends on actual delivery at the probe tip.

At low remaining volume, level detection and probe positioning determine whether sample aspiration can continue. Shallow immersion can draw air, while deep immersion increases external wetted area; wall contact and tip droplets also alter delivery. If aspiration lags, check inlet resistance and sample condition before adjusting speed or dwell time. Longer waits consume test cycle time; microclots require an abnormal-sampling procedure.

System-liquid drive also requires assessment of system-liquid mixing into the sample: a correct total weighed volume can still accompany a change in sample concentration. Observe delivered volume and sample preservation separately.

  • Sample aspiration and transfer

Further technical reading

  • Choosing a 100, 250 or 500 μL Piston Pump for Clinical Chemistry Dosing
  • Why Do Air Bubbles Delay or Bias Piston Pump Dispensing?

05Reagent dispensing: R1/R2 delivery and refill coordination

R1/R2 names, volumes and addition order must come from the method sheet for the same assay. Reusable reagent probes can aspirate from individual bottles and dispense into cuvettes, as described in the AU5800 manual. For this arrangement, check internal and external probe washing after each aspiration rather than assuming a permanently reagent-filled supply line.[1]

For dedicated reagent branches using one aspiration for multiple dispenses, subtract the program's priming and reserve volumes from available capacity before scheduling refills. Independent channels reduce switching but add components, space and retained liquid. Shared channels reduce hardware but add flushing consumption and waiting. After bottle changes or refills, restore filling and confirm the first dose before normal dispensing.

  • Metered reagent dispensing

Further technical reading

  • How Should Piston Pump Aspiration Speed and Dwell Be Set for Viscous Reagents?

06Fluidic switching: valve positions, shared volume and flushing

Define the source, destination, pump action and allowable pressure for each valve position. Shared segments retain the previous liquid after the valve reaches position. Geometric volume can estimate an initial flushing volume; adequate cleaning is established by residue and assay results under the worst-case switching sequence.

When returning from cleaner to system liquid, flush to waste before resuming metering. If certain reagent combinations require long flushing, independent channels may be preferable to additional rinsing. Compare cross-effects, per-test liquid consumption and cycle time together.

  • Sample, reagent and wash-path switching

Further technical reading

  • Rotary Valve Selection Guide: From Fluid Paths to System Validation

07Probe and cuvette washing: treat inner surfaces, outer surfaces and cuvette bottoms separately

Probe interiors require adequate flushing and effective wash solution; exteriors require the wash well to cover the wetted area; cuvette bottoms depend on aspiration position and waste capacity. For insufficient supply, examine installed resistance and concurrent demand. For poor coverage or misplaced aspiration, correct positioning and coverage before assessing whether increased supply helps.

Use separate sequences for sample carryover and reagent cross-effects. Follow high samples with low samples to examine sample residue; choose potentially interacting assay sequences for reagents. The AU5800 manual describes internal/external reagent-probe washing separately from contamination-avoidance parameters, showing that assay-combination controls complement routine washing.[1]

  • Probe/cuvette wash supply and waste removal

Further technical reading

  • What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?
  • Why do multiple wash nozzles deliver different flows from one miniature diaphragm pump?

08Tubing and fittings: why recheck shared volume after servicing?

Changes in tube length, bore or fitting cavities alter retained volume, resistance and arrival delay together. Shorter paths reduce residue, but retain bend radius and stress margins needed for assembly and maintenance. Material selection must cover every wetted surface of tubing, seals, valves and probes, checked against actual reagents, cleaners and contact durations.

Record assembly dimensions and component versions during maintenance. After replacement, check leaks, prime and purge air, then compare routine doses and first-cuvette results. Leak-tight interfaces alone do not address effects from changed shared volume.

  • Tubing and fittings in clinical chemistry analyzers

09Level, bubble and pressure detection: linking alarms to specific tests

Level signals help establish aspiration position; pressure and bubble signals describe local fluidic conditions. Pressure changes during aspiration and dispense mean different things. Interpret anomalies using action stage, valve position and intended air segments rather than one diagnostic threshold.

The instrument controller associates signals with samples, assays and cuvettes, recording commanded volumes, available estimates and measured values separately. Without measurement, do not label the command as delivered volume. After partial dispensing, handle the cuvette and decide on retesting according to the method, avoiding repeated addition onto an unknown remaining volume.

  • Bubble and pressure detection in clinical chemistry analyzers

10Explore configurations through product application guides

Use the following routes to narrow candidates when discussing configurations with FOREACH, then provide actual doses, wetted media, allowable deviations and cycle times. Model-specific performance must be based on the corresponding specifications and operating-point data.

For reagent dispensing with multiple metering units, start with HMD multichannel configurations combining glass syringes and solenoid valves. When multiple sources share one metering unit sequentially, compare HLD configurations combining a glass syringe and multiport rotary valve. Port count denotes selectable paths; verify it separately from independent metering-channel count.[2][3]

  • Small-dose metering: compare usable stroke, wetted arrangements and actual delivery for both piston and syringe pumps. Pump category alone does not establish priority.
  • Batch dispensing: plan refills from usable capacity and individual dose, comparing how larger capacity and repeated refills affect small-dose performance and cycle time.
  • Washing and waste: select supply and air-containing aspiration configurations separately and validate simultaneous multi-station operation.
  • Shared paths: reduce residue by considering rotary valves, probes, tubing and fittings together, and compare extra flushing costs with independent channels.
  • State detection: choose configurations according to target faults, sensor positions and control programs. Confirm level functions and pressure detection separately.
  • Organize configurations using the IVD product–task matrix
  • Piston pumps in clinical chemistry analyzers
  • Syringe pumps in clinical chemistry analyzers
  • Diaphragm pumps in clinical chemistry analyzers
  • Multiport rotary valves in clinical chemistry analyzers
  • Sampling and wash probes in clinical chemistry analyzers
  • Bubble and pressure detection in clinical chemistry analyzers
  • Tubing and fittings in clinical chemistry analyzers

11Validate the complete cycle through application results

Confirm delivery at the cuvette first, then use assay results to evaluate contamination and recovery. Allowable deviations and carryover targets come from the specific project and must remain consistent with the test sequence and calculation method.

Validate the complete cycle through application results
StageWhat to measurePriority checks when abnormal
SampleActual received volume at boundary doses, sample preservation and high/low switching resultsAir aspiration, wall contact, interface mixing and inner/outer probe residue
ReagentIndividual aspirations or first/middle/last batch doses; first dose after bottle change and R1/R2 timingIncomplete filling, bubbles, retained droplets and partial dispensing
Switching and washingShared-segment residue, cuvette-bottom liquid and cross-effects between assaysFlushing sequence, station coverage and accumulated waste
Maintenance and monitoringFirst cuvette after maintenance, alarm detection/false alarms and affected-cuvette tracingAssembly changes, baseline and stage association
  • View common operating-point and recovery validation methods

12Common selection and recovery questions

Can one calibration correct both low sample volume and high carryover?

Diagnose them separately. For low cuvette delivery, first check filling, seals, probe tip and actual stroke. For abnormal high/low sequences, check inner/outer probe surfaces, shared segments and cuvette washing. Displacement calibration can change volume but cannot automatically remove residue.

Where should you check first if the first cuvette is abnormal after a bottle change or refill?

Identify bottle-by-bottle aspiration versus dedicated-branch refill, then check low level, valve position, priming and probe-tip condition. Compare actual first-dose and subsequent-dose delivery and retain the cuvette's action record. If delivered volume is unknown, follow the method's handling rules rather than directly resending the same dispense command.

References

View references (3)
  1. [1]Beckman Coulter — AU5800 Chemistry Analyzer Instructions for Use,B81389AA,Sample Transfer / Reagent Transfer
  2. [2]FOREACH | HMD solenoid-valve syringe pump series
  3. [3]FOREACH | HLD rotary-valve syringe pump series

Related guides

  • Clinical chemistry sample metering in detail
  • Clinical chemistry reagent dispensing in detail

Related Products

EA Standard piston pumps

EA Standard piston pumps

HLD Rotary-valve syringe pumps

HLD Rotary-valve syringe pumps

DPL30 Liquid diaphragm pumps

DPL30 Liquid diaphragm pumps

MRV3 Rotary valves

MRV3 Rotary valves

Related Technical Articles

Choosing a 100, 250 or 500 μL Piston Pump for Clinical Chemistry Dosing

Compare dose volume, stroke utilization, dispensing count and refill timing, then assess Foreach EA, SM and TM configurations using the actual clinical chemistry fluid path.

2026-09-08View Details

Why Do Air Bubbles Delay or Bias Piston Pump Dispensing?

A completed piston stroke does not prove that the intended liquid volume reached the vessel. Bubble compression temporarily stores displacement; expansion can then cause a trailing dispense or affect the next cycle. Locate where bubbles first appear and compare synchronized pressure and individual dispense masses before and after priming.

2026-09-10View Details

How Should Piston Pump Aspiration Speed and Dwell Be Set for Viscous Reagents?

A stable aspiration window depends on resistance, filling time and pressure. Test aspiration speed and the subsequent dwell independently, then select a faster combination that satisfies bias, repeatability, bubbles and cycle-time requirements. Water settings do not automatically transfer to the reagent, and completed motion does not prove fluid equilibration.

2026-09-10View Details

Rotary Valve Selection Guide: From Fluid Paths to System Validation

Map the connections required at each step before comparing specifications. Use the actual liquids, tubing and switching sequence to validate suitability for your instrument.

2026-09-30View Details