Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control SeriesTubing Assemblies
Applications
All applicationsIVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • Tubing Assemblies
  • All applications
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: [email protected]

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Applications/In Vitro Diagnostics/Hematology analyzer
< Back

Hematology analyzer fluidics: sampling, dilution and counting-channel recovery

Application guide

Hematology analyzers distribute mixed, anticoagulated whole blood to different detection channels, using dilution, lysis or staining to prepare it for counting and differentiation. Reagents and processing conditions vary with each channel's detection method, so the fluidic configuration should follow its task.[1]

Sampling repeatability is only the starting point. Check within the same cycle whether sample and diluent are delivered in the correct ratio, cells enter the detection zone uniformly and the blank recovers before the next sample. First distinguish the metering architecture, then arrange reagent metering, detection-fluid supply, cleaning and waste removal separately.

Scope

For automated hematology analyzers that divide whole-blood samples among processing channels; predilution, closed-tube sampling and other sample modes require model-specific operating conditions.

Application guide

Hematology analyzer

Hematology analyzer
Diaphragm pumps in hematology analyzers
Hematology-channel cleaning and waste removal
Bubble and pressure sensing in hematology analyzers
Tubing and fittings in hematology analyzers
Multiport rotary valves in hematology analyzers
Detection-channel switching and fluid renewal
Piston pumps in hematology analyzers
Whole-blood sampling and delivery
Sampling probes in hematology analyzers
Syringe pumps in hematology analyzers
Diluent and reaction-reagent metering

On this page

Which branches make up a hematology analyzer fluidic system?Common product roles and application guidesOperating conditions to confirm at each stageSampling: retain mixing and low remaining sample-volume conditions in evaluationDilution and lysis: actual ratios in each counting channelDetection channels: sheath flow, apertures and fluid renewalCleaning and waste removal: restore the channel for the next sampleState identification: distinguish blockage from air aspirationExplore configurations through product application guidesValidate the complete cycle through application resultsCommon selection and recovery questionsReferences

01Which branches make up a hematology analyzer fluidic system?

Mixing and sampling→Metering and distribution→Dilution and reagent treatment→Channel counting→Cleaning and blank recovery
Liquid-handling task sequence

Arrows indicate task relationships, not that every component is connected in series in one tube. Check valve positions, pump wetted boundaries and operating states for each branch.

Typical branches handle whole-blood sampling, diluent metering, channel-reagent addition, detection-fluid supply, cleaning and waste removal. This sequence describes functional relationships; channel operations may be interleaved by the instrument scheduler.

With a metering valve, first fill the metering chamber completely, then isolate and transfer its fixed volume; filling, sealing and residue also affect delivery. With pump-displacement metering, consider pump stroke, probe tip, tubing and isolation interface as one system. Blood passing through the sampling probe does not necessarily enter the pump chamber: system-liquid drive and direct sample contact require different materials and cleaning coverage.

02Common product roles and application guides

Assign a clear task to each component before combining branches. A shear-type metering valve separates a volume using its metering chamber; an ordinary multiport rotary valve primarily selects paths. Define their functions and acceptance checks separately.

Common product roles and application guides
ProductTask in a hematology analyzer
Piston pump / syringe pumpDrive sampling or meter diluent and channel reagents; configure for direct liquid contact or system-liquid drive
Liquid diaphragm pumpLiquid replenishment and cleaning-fluid supply, combined with reservoirs, pressure stabilization or point-of-use metering
Gas–liquid diaphragm pumpRemove wash fluid and aerated waste, accommodating simultaneous inflow from multiple branches
Multiport rotary valves / solenoid valvesSelect sources, channels and cleaning paths; open and close branches
Sampling probes, tubing and fittingsProvide sampling and connections while controlling residue, inlet resistance and assembly variation
Bubble and pressure detectionProvide phase-specific abnormality signals, interpreted by the instrument with valve position, sample and action
  • Piston pumps in hematology analyzers
  • Syringe pumps in hematology analyzers
  • Diaphragm pumps in hematology analyzers
  • Multiport rotary valves in hematology analyzers
  • Sampling probes in hematology analyzers
  • Bubble and pressure sensing in hematology analyzers
  • Tubing and fittings in hematology analyzers

03Operating conditions to confirm at each stage

Provide the following information separately for each channel. A shared liquid source reduces components, but replenishment or pressure fluctuations can affect several channels simultaneously; dedicated branches require more space, tubing and maintenance points.

Operating conditions to confirm at each stage
StageConditions determining configuration
Whole-blood samplingInterval from mixing to aspiration, dose, minimum usable remaining volume, open-tube or piercing mode, liquid-contact arrangement
Dilution and channel reagentsActual dose per branch, serial-dilution sequence, mixing location and processing time
Detection-fluid supplySample and sheath-flow requirements, detection-window duration, permissible pressure and flow fluctuations
Cleaning and waste removalSurfaces requiring cleaning, cleaning agent, peak simultaneous drainage and blank-recovery requirements
Recovery from abnormalitiesSensor locations, sample association, steps that permit retry and termination conditions

04Sampling: retain mixing and low remaining sample-volume conditions in evaluation

Cell settling makes sampling time and position important to sample representativeness. Arrange mixing according to sample and assay requirements, then keep the interval from mixing completion to aspiration consistent. A mode for low remaining sample volume must account for container-bottom geometry, probe descent position and available sample volume; lowering the probe further to obtain liquid may increase liquid carried on its outer surface.

If aspiration lags, first check probe contact with the container wall, inlet resistance, bubbles and sealing, then consider reducing aspiration speed or extending the wait; extra waiting consumes cycle time. Clots or abnormal resistance should trigger the instrument's corresponding exception procedure. Simply increasing suction may carry the problem into downstream channels.

  • Whole-blood sampling and delivery

Further technical reading

  • Piston Pump Accuracy, Repeatability and Resolution Explained

05Dilution and lysis: actual ratios in each counting channel

Red-cell, white-cell and other channels have different processing objectives; lysis, staining and dilution sequences must match the specific method. First measure sample and reagent volumes delivered at the mixing location separately, then check mixing and processing time, and finally assess counting and differentiation results.

The same total volume can contain different sample proportions, so an acceptable total still requires checking each branch's contribution. Serial dilution also requires recording the preceding stage's mixing state and the position of secondary aspiration. Where sample and diluent doses differ greatly, prioritize separate metering capacities. Sharing a metering unit reduces hardware but increases flushing and sequential scheduling time.

  • Diluent and reaction-reagent metering

Further technical reading

  • What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

06Detection channels: sheath flow, apertures and fluid renewal

The detection structure determines whether sheath flow is needed. Impedance counting can also use sheath-flow focusing: the Sysmex DC sheath-flow approach guides diluted cells through the center of the aperture and then toward the drain. List sample flow, focusing flow and drainage separately rather than deciding whether sheath fluid is needed solely from the label 'impedance' or 'optical'.[2]

When a syringe pump supplies liquid during a detection window, usable capacity must cover that window and necessary auxiliary actions. If replenishment interrupts flow, schedule it between measurements. For longer continuous supply, compare reservoir-based pressure stabilization or alternating supply and check handover fluctuations. Source-supply pumps and point-of-use flow control serve different tasks; selection must account for the detection channel's actual resistance.

  • Detection-channel switching and fluid renewal

Further technical reading

  • Rotary Valve Selection Guide: From Fluid Paths to System Validation

07Cleaning and waste removal: restore the channel for the next sample

The inner and outer sampling-probe surfaces, mixing chamber, aperture or flow cell each need adequate cleaning coverage. For inadequate supply, first check installed resistance and flow splitting between branches. For delayed drainage, check the waste circuit, foam and container state. Increasing supply also increases waste-removal load; match pump capacity to peak simultaneous inflow.

After blockage, use the model's permitted flushing or backflushing procedure and check channel flow, blank and specified quality controls before resuming samples. Judge cleaning success by restored channel performance, not merely by pump operation or a cleared alarm.

  • Hematology-channel cleaning and waste removal

Further technical reading

  • What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?
  • IVD Waste Aspiration: Liquid Pump or Vacuum Pump? Direct vs Indirect Aspiration

08State identification: distinguish blockage from air aspiration

Establish separate pressure baselines for sampling, reagent addition, counting and cleaning. Inlet obstruction, an exhausted liquid source, bubbles, valve changes and leaks can all alter the trace; combine the timing of changes with bubble or level signals to narrow the cause.

The instrument control system associates sensor outputs with sample ID, current channel and step. Distinguish commanded volume, stroke-estimated volume and independently measured volume in records. If an abnormality occurs after partial aspiration or dispensing, preserve completion status before deciding to resample, remake that sample's diluted mixture or terminate the determination.

  • Bubble and pressure sensing in hematology analyzers

Further technical reading

  • Why Do Air Bubbles Delay or Bias Piston Pump Dispensing?

09Explore configurations through product application guides

For multiple metering units, compare channel configurations in the FOREACH HMD series. For one metering unit selecting multiple liquid sources sequentially, compare the HLD syringe and multiport rotary-valve combination. Multiple metering channels and multiport routing are separate dimensions; more ports do not automatically create more branches that can meter simultaneously.[3][4]

Compare piston and syringe pumps by minimum dose, auxiliary strokes, replenishment timing and wetted path. Select cleaning and waste pumps by installed flow and gas–liquid conditions. For valves, prioritize shared-path residue; for probes and fittings, cleaning coverage and dead volume. Model-specific capacity and materials must suit the actual reagents, cleaning agents and operating sequence.

  • View the IVD product–task matrix
  • Piston pumps in hematology analyzers
  • Syringe pumps in hematology analyzers
  • Diaphragm pumps in hematology analyzers
  • Multiport rotary valves in hematology analyzers
  • Sampling probes in hematology analyzers
  • Bubble and pressure sensing in hematology analyzers
  • Tubing and fittings in hematology analyzers

10Validate the complete cycle through application results

Record metering, ratios, detection flow and recovery separately, then look for relationships within the same test batch. Acceptance targets are defined by the corresponding method and project.

Validate the complete cycle through application results
StageActual measurementPriority areas to investigate
SamplingDelivered volume at normal and low remaining levels; carryover from high to low samplesMixing, air aspiration, clots and residue inside and outside the probe
Dilution processingDelivered volume by branch, mixing and processing time, assay resultsCapacity configuration, shared residual liquid and aspiration position
Detection-fluid supplyFlow, pressure, bubbles and counting blank during the detection windowReplenishment handover, focusing-fluid supply, aperture or flow cell
Cleaning recoveryPeak waste removal, blank after flushing and first sample after recoveryFlow splitting, foam, residual liquid and maintenance assembly
Abnormality identificationPhase-specific detection and false alarms, response time and sample associationBaseline differences, sensor position and control state
  • General validation methods and boundary operating conditions

11Common selection and recovery questions

Why can counts remain abnormal when sampling repeatability passes?

Continue by checking actual sample-to-reagent ratios, mixing and processing time, then flow, bubbles and channel blank during detection. Repeatability describes only the consistency of one action; persistent delivery bias or channel residue can still affect results.

Does increasing syringe-pump capacity guarantee continuous sheath-fluid supply?

Greater capacity can extend a single supply window while changing stroke utilization and replenishment scheduling. Calculate the required supply volume first, then check whether replenishment fits between measurements. If no gap is available, evaluate alternating or pressure-stabilized supply and its switching fluctuations.

References

View references (4)
  1. [1]Sysmex — Haematology technologies
  2. [2]Sysmex — DC sheath flow detection method
  3. [3]FOREACH | HMD solenoid-valve syringe pump series
  4. [4]FOREACH | HLD rotary-valve syringe pump series

Related Products

EA Standard piston pumps

EA Standard piston pumps

HLD Rotary-valve syringe pumps

HLD Rotary-valve syringe pumps

DPL30 Liquid diaphragm pumps

DPL30 Liquid diaphragm pumps

ABD Air-bubble detectors

ABD Air-bubble detectors

PDM5 Pressure sensors

PDM5 Pressure sensors

Related Technical Articles

Piston Pump Accuracy, Repeatability and Resolution Explained

Learn why theoretical volume per step is not the minimum reliable dose, and how accuracy, repeatability, valves, bubbles and working stroke affect dispensing results.

2026-09-06View Details

What Is a Programmable Syringe Pump? Operation, Fluid Routing and Automation

A programmable syringe pump uses a motor to control syringe plunger displacement and speed, executing aspiration, dispensing, pauses and repeated actions through commands. With a routing valve, it can connect reagent reservoirs, reaction vessels and wash paths into an automated dispensing process. Instrument development therefore needs to define volume, fluid connections and timing together.

2026-09-28View Details

Rotary Valve Selection Guide: From Fluid Paths to System Validation

Map the connections required at each step before comparing specifications. Use the actual liquids, tubing and switching sequence to validate suitability for your instrument.

2026-09-30View Details

What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?

Cleaning pump, wash pump and rinse pump usually name different duties within an IVD cleaning sequence, not three fixed pump mechanisms. Selection starts with wash volume, cycle rate, line resistance, wetted materials, and whether the duty is liquid supply or waste aspiration.

2026-09-03View Details

Discuss Your IVD Fluidic Requirements

Share the instrument type, sample and reagent properties, working volumes, cycle timing, fluid path and acceptance criteria so we can discuss suitable components and prototype evaluation.

Discuss your application